| Model Number |
56.54.27P |
| Medical Device Problem Code |
Appropriate Device Problem Term/Code Not Available (3191)
|
| Health Effect - Clinical Code |
Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
02/06/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The user facility in japan reported after surgery, the tip of the probe became curled solid, and could not be pulled out of the incision.An incision was required to pull the probe out of the patients eye.No health problems were observed in the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
This investigation is on going.
|
| |
|
Additional Manufacturer Narrative
|
|
This investigation is complete.
|
| |
|
Additional Manufacturer Narrative
|
|
We received approximately 1-1/2 inches of fiber from the distal end of the probe wrapped in bubble wrap and affixed to a piece of cardboard.This packaging was inside the original, opened product pouch.The cannula and valve were present on the fiber.Visual examination under magnification shows residual surgical matter on the fiber and cannula.The tip of the fiber was no longer conical.The current state has the shape of an irregular, roughly spherical, ball.The diameter of fiber at the distal tip exceeds the od of the cannula.Accumulated debris on the end of the light probe resulted in the concentration of illumination energy exceeding the glass transition temperature of the fiber material.The sterilization and lot history records were reviewed and found to be acceptable.The trend analysis, risk analysis and/or directions for use review were considered acceptable, with the product performing within anticipated rates.No corrective action required.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction: d4 corrected udi number.
|
| |
|