• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GIVEN IMAGING LTD., YOQNEAM PILLCAM; SYSTEM, IMAGING, GASTROINTESTINAL, WIRELESS, C

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GIVEN IMAGING LTD., YOQNEAM PILLCAM; SYSTEM, IMAGING, GASTROINTESTINAL, WIRELESS, C Back to Search Results
Model Number FGS-0667
Medical Device Problem Code Entrapment of Device (1212)
Health Effect - Clinical Codes Obstruction/Occlusion (2422); Unintended Radiation Exposure (4565)
Date of Event 02/07/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, the capsule was ingested and after a week the patient developed obstructive symptoms.Patient had an x-ray and showed the patency capsule was still in the small bowel and intact.Patient was admitted to the hospital and had surgery laparoscopy moved to laparotomy for lysis adhesion, small bowel resection for obstruction and appendix removal.Patient was currently in the hospital for post surgery.
 
Additional Manufacturer Narrative
Additional information: g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was not returned, but two x-ray photos were ava ilable for evaluation.Visual inspection noted the capsule to be in the pelvis and it may be in the sigmoid colon or rectum.It was reported that the capsule remained in the patient longer than expected.The reported issue was confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.The instructions included with this device provide the following guidance: patients with suspected or known intestinal strictures may be at increased risk for bowel obstruction due to impaction of the intact patency capsule at the site of the narrowing or stricture, particularly during the first 1-2 days after capsule administration.Should this occur, patients may require bowel decompression until the capsule disintegrates or surgical intervention, if symptoms persist.Patients should be made aware of the signs and symptoms of bowel obstruction (abdominal pain, nausea, vomiting, and distention) and be advised to contact their physician should these symptoms occur anytime between capsule ingestion and documented excretion.Patients should also be informed that surgical intervention may be required if obstruction were to occur.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PILLCAM
Common Device Name
SYSTEM, IMAGING, GASTROINTESTINAL, WIRELESS, C
Manufacturer (Section D)
GIVEN IMAGING LTD., YOQNEAM
yetsira 13 street
yoqneam 20692
IS  20692
Manufacturer (Section G)
GIVEN IMAGING LTD., YOQNEAM
yetsira 13 street
yoqneam 20692
IS   20692
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key16491534
Report Number9710107-2023-00036
Device Sequence Number6787253
Product Code NEZ
UDI-Device Identifier07290110360818
UDI-Public07290110360818
Combination Product (Y/N)N
Initial Reporter StateKY
Initial Reporter CountryUS
PMA/510(K) Number
K090557
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/31/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/07/2023
Device Model NumberFGS-0667
Device Catalogue NumberFGS-0667
Device Lot Number57832
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/16/2023
Supplement Date Received by Manufacturer03/29/2023
Initial Report FDA Received Date03/06/2023
Supplement Report FDA Received Date03/31/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/07/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age41 YR
Patient SexFemale
-
-