STRYKER TRAUMA KIEL TROCHANTERIC NAIL KIT, TI GAMMA3® Ø11X200MM X 125°; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Model Number 3125-1200S
Medical Device Problem Codes Break (1069); Fracture (1260); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Codes Failure of Implant (1924); Implant Pain (4561)
Date of Event 12/13/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Once the investigation has been completed any additional information will be reported in a supplemental report.
 
Event or Problem Description
As reported: "pain in the operated area without trauma with significant increase in the course.X-ray diagnosis revealed sintering of the gamma nail with fracture of the distal locking screw.Patient was operated again.".
 
Additional Manufacturer Narrative
Please note correction to h6 (device code).The reported event could be confirmed since the device was returned and matches the alleged failure.Device inspection revealed the following: the received nail is completely broken in the webs of the proximal lag screw hole.Heavy explantation marks were observed on the inner surface of the hole and the fracture surface as well, mostly in the distal segment.We can see the load-bearing points on the proximal part of the broken nail which indicates overloading.We can also see the presence of breakage in an oblique manner which indicates towards application of high torsional load.Bearing points of the lag screw at the medial edge of the proximal hole showed significant deformation, indicating the high compressive load.Signs of overloading were also evident at the distal oblong hole.The breakage pattern resembles a fatigue breakage, having equal portions of fatigue zone and instantaneous zone indicating that the breakage was initiated in a fatigued manner but quickly broke instantaneously under high compressive loading.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.Based on the investigation, the root cause was attributed to patient related factors as breakage was due to high compressive load and with available overloading signs on nail.If any additional information is provided, the investigation will be reassessed.
 
Event or Problem Description
As reported: "pain in the operated area without trauma with significant increase in the course.X-ray diagnosis revealed sintering of the gamma nail with fracture of the distal locking screw.Patient was operated again.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TROCHANTERIC NAIL KIT, TI GAMMA3® Ø11X200MM X 125°
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM  D-24232
Manufacturer (Section G)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM   D-24232
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key16491699
Report Number0009610622-2023-00075
Device Sequence Number5660030
Product Code HSB
UDI-Device Identifier04546540375087
UDI-Public04546540375087
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K034002
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/01/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date09/30/2024
Device Model Number3125-1200S
Device Catalogue Number31251200S
Device Lot NumberK038623
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/10/2023
Supplement Date Received by Manufacturer05/08/2023
Initial Report FDA Received Date03/06/2023
Supplement Report FDA Received Date06/01/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/14/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age88 YR
Patient SexFemale
-
-