| Model Number |
IPN922900 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/08/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Qn#(b)(4).The complaint sample was received at the manufacturing site for evaluation.Prior to review of the sample there was no infected tag on the complaint sample.The product pouch was attached.The lot number and size were same as the complaint description.Proper investigation cannot be performed as the device still inside the complaint packaging and no infected tag to confirm the product has been disinfected.From the outer package reviewed, the complaint description of "cuff separated from tube during use" is confirmed as the part was observed as separated (cuff and airway tube).The device history for the packaging lot 11f22do273 was reviewed and found no abnormalities with the complaint lot.The device history review for the assembly process lot affected was reviewed and there was no nonconformances issued for the assembly lot.It was concluded that the complaint root cause is undetermined/unknown.Further investigation will be completed after clarity of the complaint sample status is obtained from complaint handling coordinator and a follow-up report will be submitted when the sample investigation is completed.
|
| |
|
Event or Problem Description
|
|
Reported issue: patient was under general anesthesia for a urological procedure.When waking up, the nurse was removing the laryngeal mask and at the same time the patient clenched his teeth and the tube separated from the cuff.The inurse had to extract the cuff manually.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).The complaint sample was received and a visual inspection was performed on the product.The lot number and the size was the same as the complaint description.After verifying the product the actual defect was the airway tube was detached from the backplate.The inflation line also was detached from the cuff.On visual inspection of the airway tube and backplate it was observed that the awt-backplate joint area, showed the airway tube area has glue mark on the airway tube.In the device history review for the packaging lot 11f22do273 no abnormalities were found.In the device history review for the assembly process for the affected lot no nonconformance were issued.Since there was the presence of glue on the airway tube area and airway tube was detached due to patient clenching his teeth the root cause identifies as undetermined/unknown.There was limited information related to how the patient clenched the part or how doctor/user remove the device.In conclusion, based on the complaint description of "cuff separated from tube during use", reviewing of the device history of the packaging and assembly lot, there was no abnormality found.After reviewing the sample received, the defect was the same as the complaint description, the airway tube was detached from the backplate.Since there was glue marking present on the airway tube area and the airway tube was detached due to the patient clenching his teeth, the root cause identifies as undetermined/unknown.
|
| |
|
Event or Problem Description
|
|
Reported issue: patient was under general anesthesia for a urological procedure.When waking up, the nurse was removing the laryngeal mask and at the same time the patient clenched his teeth and the tube separated from the cuff.The inurse had to extract the cuff manually.
|
| |
|