Model Number BB811
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/06/2023
Type of Reportable Event Malfunction
Event or Problem Description
Medtronic received information that prior to use of a fusion oxygenator, the customer was inspecting the device prior to use.During setup and before priming, the tma port was visibly inspected and when the temperature probe atp210 was gently inserted the port came away from the body of the device with ease.The oxygenator was cut out of the custom pack and replaced.There was no patient involvement, so no adverse effect occurred.It was also reported that no force was applied and there was no damage to any packaging.
 
Additional Manufacturer Narrative
Device evaluation summary: visual inspection shows the tma is loose and can be easily removed.Additional visual inspection shows evidence of adhesive residue.Device appears not to have been used.Reason for return was confirmed.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Conclusion: the complaint was confirmed for the fusion oxygenator's loose temperature monitoring adapter (tma).The issue was verified via analysis of the returned device.Review of this unit¿s device history record found no abnormalities or non-conforming material reports (ncmrs) initiated during manufacturing that would cause or contribute to the reported event.This unit passed all required testing and inspections during manufacturing.Customer communication of this issue has been initiated via fa1302, which provides device use recommendations for affected product.Root cause investigation concluded the presence of ebs (ethylene bis-stearamide) wax on the new tma component affected the material properties that caused the torque specification to not be met.There were no patient/clinical safety issues reported.Trends for issues with this product are reviewed at quarterly quality meetings.Correction d4.5 (unique identifier (udi) #): this field has been updated.Correction h9: the number in this field has been updated to the z number.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FUSION HOLLOW FIBER OXYGENATOR
Common Device Name
OXYGENATOR, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
PERFUSION SYSTEMS
7611 northland dr
brooklyn park MN 55428
Manufacturer (Section G)
PERFUSION SYSTEMS
7611 northland dr
brooklyn park MN 55428
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key16494009
Report Number2184009-2023-00112
Device Sequence Number12357907
Product Code DTZ
UDI-Device Identifier00643169178168
UDI-Public00643169178168
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K203111
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Remedial Action Notification,Recall
Type of Report Initial,Followup
Report Date (Section B) 03/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2024
Device Model NumberBB811
Device Catalogue NumberBB811
Device Lot Number224986839
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/01/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/10/2023
Supplement Date Received by Manufacturer03/15/2023
Initial Report FDA Received Date03/07/2023
Supplement Report FDA Received Date03/28/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/26/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-1023-2023
Patient Sequence Number1
Patient SexUnknown
-
-