| Catalog Number |
611.74 |
| Medical Device Problem Code |
Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/31/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A physician reported that after a cataract surgery, it was observed under a microscope that an ophthalmic retractor stopper part was torn.The damaged part was a circular silicone part, extraocular.The retractor was removed at the end of the surgery and the surgery was completed without any problem.There was no patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record (dhr) traceable to the reported lot number indicates that the product was processed and released according to the product¿s acceptance criteria.The sample received with inner blister, outer blister and cover foil showing the lot information.The sample shows macroscopic signs of damage.The silicone stopper was missing.The sample shows no signs of surgery residues.The sample was visually inspected with the aid of a photomicroscope with various magnifications.The sample was visually inspected.It was noticed that the silicone stopper was missing.As the blister was opened by the customer, he handled the product.The point in time when the malfunction occurred, can no longer be determent.It cannot be determined how or where the damage to the sample occurred; therefore a root cause for the customers reported event could not be determined.However, the damage was consistent with damage that can occur due to improper handling.The exact root cause for the customers reported event is unknown, therefore, specific action cannot be taken.Complaints are reviewed and monitored at regular intervals for adverse trends.No adverse trends have been observed associated with the reported product and event.No action has been identified for this reported event.The manufacturer internal reference number is: (b)(4).
|
| |
|
Search Alerts/Recalls
|