|
Qn#(b)(4).The customer returned one cartridge of 30130 vclude ti sw 6/cart 30/box and one 70362 vclude aplr sw cvd 6" 152mm applier for investigation.The cartridge and applier were examined with and without magnification.Visual examination of the cartridge revealed obvious signs of use in the form of biological material.There were four clips remaining in the cartridge.The applier returned did not appear to be misaligned.Functional inspection was performed on the four remaining clips using the returned appliers.All four clips from the returned cartridge was able to load properly into the returned appliers and was able to be applied to over-stressed surgical tubing.The reported defect could not be replicated based upon the sample returned.The customer did not provide a lot number; therefore, a device history record review was performed based upon a lot number from the sales history data of the customer.No relevant findings were identified.The ifu for this product states: "remove the applier from cartridge and verify that the clip legs do not protrude beyond the end of the jaws.The clip should be seated firmly." "avoid premature closure of the applier jaws which can cause a clip to partly close and slip back." "user to establish unobstructed application site visibility and confirm location of clip placement prior to applying clip.User should not twist applier on vessel to gain visibility." "at the surgical site the surgeon should close the applier ring handles with a smooth, firm, continuous motion until the clip is fully closed.Releasing the pressure on the ring handles will cause the applier jaws to spring open." "appliers are precision instruments and must be handled with care.Inspect for wear or damage and confirm adequate clip pick-up prior to use.Return to vesocclude medical, llc, llc for repair as necessary.If not done a patient injury could occur." "always check the alignment of applier jaws before use.When closed, jaw tips should be directly aligned and not o ffset.Alignment of the jaw is critical for safe application of the clip.If this is not done, patient injury could occur." "inspect instruments after each sterilization cycle and prior to each use as they may be damaged during transit to the customer, during rece iving at the customer's site, during use in a previous procedure or during the cleaning or sterilization processes.All moving parts must be inspected for wear and confirmed to be functional.Confirm smooth operation during opening and closing of instrument handle and reinforcing screw." "inspect instruments for rust, pitting, cracking or burrs, staining or discoloration, and worn or broken parts.Repair or replace any instrument found not to be acceptable." "do not use an instrument with broken, cracked or worn parts." the customer reported complaint could not be confirmed.One cartridge and one applier were returned.Visual examination revealed obvious signs of use in the form of biological material.There were four clips remaining in the cartridge.A dhr review was performed on the potential lot number of the returned cartridge and no relevant findings were identified.Functional inspection was performed on the four remaining clips using the returned applier.All four clips from the returned cartridge was able to load properly into the returned applier and was able to be applied to over-stressed surgical tubing.The ifu for this product states "always check the alignment of applier jaws before use.When closed, jaw tips should be directly aligned and not offset.Alignment of the jaw is critical for safe application of the clip.If this is not done, patient injury could occur." "appliers are precision instruments and must be handled with care.Inspect for wear or damage and confirm adequate clip pick-up prior to use.Return to vesocclude medical, llc, llc for repair as necessary.If not done a patient injury could occur." teleflex will continue to monitor and trend on complaints of this nature.
|