MAUDE Adverse Event Report: ONKOS SURGICAL INC. ELEOS; HYBRID CEMENTED SEGMENTAL STEM, 13MM X 120MM, STRAIGHT, FLUTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ONKOS SURGICAL INC. ELEOS; HYBRID CEMENTED SEGMENTAL STEM, 13MM X 120MM, STRAIGHT, FLUTED Back to Search Results
Model Number HC-13120-03M
Medical Device Problem Codes Use of Device Problem (1670); Improper or Incorrect Procedure or Method (2017)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/25/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during an eleos proximal femoral replacement surgery on (b)(6) 2022, the surgeon implanted the eleos modular collar cemented stem without a eleos modular collar or eleos collar locking ring.This type of stem is intended to be implanted with a modular collar and collar locking ring.No adverse events have been reported from this malfunction.This event will be reportable to the fda as a malfunction.
 
Additional Manufacturer Narrative
The investigation is in process.When the investigation is complete, a supplemental mdr will be submitted accordingly.
 
Additional Manufacturer Narrative
It was reported that during an eleos proximal femoral replacement surgery on 2(b)(6) 2022, the surgeon implanted the eleos modular collar cemented stem without a eleos modular collar or eleos collar locking ring.This type of stem is intended to be implanted with a modular collar and collar locking ring.No adverse events have been reported from this malfunction.Based upon the device history records, sterilization records, and interview with the sales representative the investigation suggests that the root causes of the issue is not likely to be related to the manufacture, and/or sterilization of the component.The surgical technique specifies the eleos modular collar cemented stem to be implanted with the eleos modular collar and eleos collar locking ring.The surgical technique for the eleos proximal femoral replacement (elss pf 06.09.21 v4) was reviewed and the following applicable excerpts were found: modular collar and stem implant assembly: o assemble the collar over the intermedullary shaft of the hybrid stem, starting distally and sliding up the stem proximally.Confirm the taper of the collar threads matches the taper of the stem shaft.O assemble the locking ring over the proximal end of the hybrid stem in preparation for threading to the collar.O hand-tighten the locking ring while holding the collar in a firm fingertip grip.O insert the counter-torque socket into the assembly tower base hole.O align the t-shaped anti-rotation boss of the stem with the matching t-slot of the countertorque socket.Insert the hand-tightened stem collar-locking ring assembly into the countertorque socket.O slide the open socket of the torque wrench on the hex of the locking ring ensuring the handle of the torque wrench is fully seated.O insert the counter-torque handle into the indicated hole of the counter-torque socket facing the user.O in a scissor motion, apply force on the torque wrench handle in the direction of the "tightening" arrow on the upper surface of the counter-torque socket until the handle shaft reaches a minimum of 35nm on the torque wrench scale indicator.O remove the torque wrench and stem-collar locking ring assembly from the socket.O rotate the stem-collar locking ring assembly until the locking ring impactor hole is aligned vertically up.O insert the stem-collar-locking ring assembly in the hex socket of the counter-torque socket, ensuring the peening hole is visible through the guide ring of the counter-torque socket.O thread a single-use impactor tip on the collar impactor.O insert the impactor tip into the guide ring of the counter-torque socket, ensuring that the black ring on the tip is no longer visible (it is covered by the guide ring).O while firmly holding the collar impactor, firmly strike the impactor with the surgical mallet five times.O remove the stem-collar-locking ring assembly from the hex socket of the counter-torque socket.Proceed to the standard surgical technique step for standard impaction to the implant and/or optional midsections the root cause of this issue was not able to be determined conclusively.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELEOS
Common Device Name
HYBRID CEMENTED SEGMENTAL STEM, 13MM X 120MM, STRAIGHT, FLUTED
Manufacturer (Section D)
ONKOS SURGICAL INC.
77 e halsey road
parsippany NJ 07054
Manufacturer Contact
upasana basnet
77 e halsey rd
parsippany, NJ 07054
MDR Report Key16497048
Report Number3013450937-2023-00041
Device Sequence Number17734024
Product Code KRO
UDI-Device IdentifierB278HC1312003M0
UDI-PublicB278HC1312003M0
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K203815
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberHC-13120-03M
Device Catalogue NumberHC-13120-03M
Device Lot Number84549-012
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/09/2023
Supplement Date Received by Manufacturer05/10/2023
Initial Report FDA Received Date03/07/2023
Supplement Report FDA Received Date06/09/2023
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/25/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age67 YR
Patient SexFemale
-
-