| Model Number |
PROPAQ MD |
| Medical Device Problem Code |
Grounding Malfunction (1271)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has not received the device for evaluation and this complaint is still under investigation.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during a routine shift check by a clinician, the device failed leakage current test.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental medwatch report is reporting the evaluation of the device and correcting information submitted on the initial medwatch report.Evaluation: the customer was contacted for return of the suspect product.The product has not been returned to zoll for evaluation.
|
| |
|