TELEFLEX MEDICAL EZ-IO POWER DRIVER; NEEDLE, HYPODERMIC, SINGLE LU Back to Search Results
Model Number IPN915667
Medical Device Problem Code Complete Loss of Power (4015)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/30/2022
Type of Reportable Event Malfunction
Event or Problem Description
Reported issue: during cpr the driver had a power failure while attempting to insert the io in the left tibia with a 45mm needle.The drill sounded charged and had a green light 2 days prior.
 
Additional Manufacturer Narrative
(b)(4).The device has not been returned for investigation.Teleflex will continue to monitor and trend related events.
 
Event or Problem Description
Reported issue: during cpr the driver had a power failure while attempting to insert the io in the left tibia with a 45mm needle.The drill sounded charged and had a green light 2 days prior.
 
Additional Manufacturer Narrative
Qn#(b)(4).The manufacturing device history file was reviewed based on the potential sn number revealed from sales history.No recorded results of manufacturing issues or anomalies were reported and passed all the release criteria.The device was released in 08/2022 and is approximately 0.5 years old.Two ifus will be referenced for this investigation.The ez-io driver ifu (8048a) states, "as with any emergency medical device carrying a backup is strongly advised protocol", "ez-io power driver led will blink red when the trigger is activated and has only 10% of battery life remaining", and "purchase and replace the ez-io power driver when the red led begins blinking".The ez-io needle ifu (8087a) states, "squeeze trigger and apply gentle, steady pressure.Important: do not use excessive force.Note: if ez-io power driver stalls and ez-io needle set will not penetrate the bone, operator may be applying too much downward pressure to penetrate bone." corrective action is not required at this time as the probable cause could not be determined based upon the information provided and without a sample.The complaint could not be confirmed as no sample was returned for analysis.A device history record review was performed based on a potential serial number from sales history, and no relevant findings were identified.Without the device to evaluate, the complaint could not be confirmed and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.
 
Event or Problem Description
Reported issue: during cpr the driver had a power failure while attempting to insert the io in the left tibia with a 45mm needle.The drill sounded charged and had a green light 2 days prior.
 
Additional Manufacturer Narrative
(b)(4).Ez-io driver 9058 (sn (b)(6)) was returned for evaluation.Upon receipt, the driver was visually inspected.When the trigger was pulled, the driver was operable and a solid green led light was displaying.The driver was then subjected to simulated sawbone insertions per driver ifu.This functional test is used to determine whether the returned device still has the capability to power a 45mm ez-io needle set into a medium and/or hard bone.The 45mm ez-io needle set is used because it represents the worst case scenario for io insertion.Two (2) sawbones with medium (50 lbs/ft^3) and hard (102 lbs/ft^3) hardness profiles with 3mm cortexes were used for the test.Each insertion was timed with a stopwatch while applying approximately 8 lbs.Of force on a weight scale.Approximately 5 to 8 lbs.Of force is necessary to penetrate bone.The driver successfully negotiated both the medium and hard saw bone insertion testing.The complaint cannot be confirmed.The driver was opened to test and record operating characteristics.Battery voltage measured as 17.84 dc volts without being activated.Battery voltage measured as 14.94 dc volts when button was activated and voltage reached a stable level.Stable defined as when whole numbers (e.G., 6 volts, 7 volts, etc.) stop changing (ignoring numbers after the decimal point).The measured voltage is slightly lower than the nominal specification of 18.2v for the battery pack upon release.This is expected after device use.The eeprom was parsed and the results reveal the charge count was 3,522,929and the cycle count was 489.The cycle count of 489 (trigger activations) is significant as it substantiates driver usage with considerations to bone density, average insertion time, storage, and frequency of testing.The eeprom results confirm that the battery is not depleted as the charge count has not exceeded the maximum depletion value of 15,300,000 per ver0190 accel biotech firmware verification test report.Additionally, 2 stall conditions were recorded.The ez-io needle ifu states, "squeeze trigger and apply gentle, steady pressure.Important: do not use excessive force.Note: if ez-io power driver stalls and ez-io needle set will not penetrate the bone, operator may be applying too much downward pressure to penetrate bone." testing confirms that there was no fault found with the returned driver.A copy of the manufacturing dhr file was reviewed.No recorded results of manufacturing issues or anomalies were reported.Two ifus w ill be referenced for this investigation.The ez-io driver ifu states, "as with any emergency medical device carrying a backup is strongly advised protocol", "ez-io power driver led will blink red when the trigger is activated and has only 10% of battery life remaining".The ez-io needle ifu states, "squeeze trigger and apply gentle, steady pressure.Important: do not use excessive force.Note: if ez-io power driver stalls and ez-io needle set will not penetrate the bone, operator may be applying too much downward pressure to penetrate bone." a review of the device history record found that the driver passed all the release criteria.The device was released in 05/2017 and is approximately 6 years old.The complaint cannot be confirmed.Functional testing has confirmed no fault found with this driver.No corrective/preventative actions will be assigned.The complaint cannot be confirmed.Functional testing revealed no fault found with this driver as the device was able to negotiate both the medium and hard saw bones during insertion testing.No defects or anomalies were observed with the returned driver.The eeprom results confirm that the battery is not depleted as the charge count has not exceeded the maximum depletion value of 15,300,000 per accel biotech firmware verification test report.Additionally, 2 stall conditions were recorded.The ez-io needle ifu states, "squeeze trigger and apply gentle, steady pressure.Important: do not use excessive force.Note: if ez-io power driver stalls and ez-io needle set will not penetrate the bone, operator may be applying too much downward pressure to penetrate bone." a device history record review was performed and no recorded results of manufacturing issues or anomalies were reported.Teleflex will continue to monitor and trend on complaints of this nature.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EZ-IO POWER DRIVER
Common Device Name
NEEDLE, HYPODERMIC, SINGLE LU
Manufacturer (Section D)
TELEFLEX MEDICAL
athlone
Manufacturer (Section G)
TELEFLEX MEDICAL
3015 carrington mill blvd
4 kamunting industrial estate
morrisville NC 27560
Manufacturer Contact
bryanna connelly
3015 carrington mill blvd
morrisville 27560
9194332672
MDR Report Key16497411
Report Number8040412-2023-00100
Device Sequence Number6240036
Product Code FMI
UDI-Device Identifier10816000013813
UDI-Public10816000013813
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
K180395
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberIPN915667
Device Catalogue Number9058
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/02/2022
Supplement Date Received by Manufacturer03/22/2023
04/20/2023
Initial Report FDA Received Date03/07/2023
Supplement Report FDA Received Date03/27/2023
04/20/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
NONE REPORTED; NONE REPORTED; NONE REPORTED
Patient SexUnknown
-
-