Model Number 453943
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/10/2023
Type of Reportable Event Malfunction
Event or Problem Description
There was a highly calcified lesion where a competitor stent would not cross.Then the orsiro mission drug-eluting stent system was selected which crossed the lesion, but not fully.The orsiro mission was withdrawn through the guide, and it was noticed that the stent was not on the balloon, but in the body.The stent was retrieved.The intervention was completed and the patient was discharged with no issues.
 
Additional Manufacturer Narrative
Combination product: yes.
 
Additional Manufacturer Narrative
Combination product: yes.The returned product was subjected to a detailed technical analysis and the corresponding product release documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for this event.The technical investigation revealed that the balloon is slightly unfolded.In addition, contrast medium residue was observed in the balloon- and inflation lumen which is indicative for the application of negative pressure.Stent imprints were observed on the exposed balloon surface, indicating that the stent was properly crimped in between the two radiopaque markers at the time of delivery.Furthermore, the hypotube was found kinked at several locations over its entire length.The stent was returned separately.The stent is severely deformed over its entire length.Review of the product release documentation confirmed that the instrument was manufactured according to specifications and passed all in-process and final inspections.As a part of final inspection every stent system undergoes visual inspection to ensure correct embedding and homogeneous crimping of the stent.Further, the stent outer diameter and the stent retention force of a defined amount of samples is tested from every lot.Based on the conducted investigations of the device being subject to this complaint, no material or manufacturing related root cause was determined.The most probable root cause for the complaint event is related to the handling during the procedure, i.E.Application of negative pressure prior to the placement of the stent across the target lesion which is warned against in the ifu.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ORSIRO MISSION (US) 2.25/18
Common Device Name
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
BIOTRONIK AG, BUELACH, SWITZERLAND
ackerstrasse 6
buelach CH-81 80
CH  CH-8180
Manufacturer Contact
6024 jean road
lake oswego, OR 97035
8772459800
MDR Report Key16497470
Report Number1028232-2023-01176
Device Sequence Number6451854
Product Code NIQ
UDI-Device Identifier07640130455993
UDI-Public(01)07640130455993(17)2307
Combination Product (Y/N)Y
Initial Reporter CountryUS
PMA/510(K) Number
P170030
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/15/2023
Device Model Number453943
Device Catalogue NumberSEE MODEL NO.
Device Lot Number05214847
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/14/2023
Supplement Date Received by Manufacturer04/06/2023
Initial Report FDA Received Date03/07/2023
Supplement Report FDA Received Date04/11/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/15/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;