| Model Number |
434891 |
| Medical Device Problem Code |
Lack of Effect (4065)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
02/15/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Two pk papyrus covered stents were used to try to seal a type iii perforation in the proximal lad, but were unsuccessful.The other stent is filed separately.
|
| |
|
Additional Manufacturer Narrative
|
|
The affected pk papyrus stent systems were not returned to biotronik and could therefore not be subjected to a technical investigation.Images of the case such as angiographies could also not be obtained.The provided clinical information (pk papyrus device registration form) and the product release documentation was reviewed to establish whether a device deficiency contributed to this event.Review of the product release documentation confirmed that the devices were manufactured according to specifications and fulfilled all the requirements of in-process and final inspection.Based on the provided documentation and the conducted review no manufacturing related root cause could be identified.The treating physician rated the occurrence as non-procedure- but device-related complication.Pk papyrus is indicated for the treatment of acute perforations of native coronary arteries and coronary bypass grafts in vessels 2.5 to 5.0 mm in diameter.As such, this device is typically used in emergent response to a life-threatening adverse procedural event unrelated to use of the pk papyrus.
|
| |
|
Search Alerts/Recalls
|