• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: THE SPECTRANETICS CORPORATION SPECTRANETICS LEAD LOCKING DEVICE; STYLET, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

THE SPECTRANETICS CORPORATION SPECTRANETICS LEAD LOCKING DEVICE; STYLET, CATHETER Back to Search Results
Model Number 518-019
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/07/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Patient's date of birth unk.Patient's weight unk.Other relevant history unk.Device lot number, expiration date unk.The device was discarded, thus no investigation could be completed.Device manufacture date unk because lot number unk.Although device damage is a known risk of complication with use of the lld, the cause for the lld breaking could not be established.Submission of this report does not, in itself, represent a conclusion by the manufacturer and/or authorized representative or the national competent authority that the content of this report is complete or accurate, that the medical device(s) listed failed in any manner and/or that the medical device(s) caused or contributed to an alleged death or deterioration in the state of the health of any person.
 
Event or Problem Description
A lead extraction procedure commenced to remove a right ventricular (rv) lead due to malfunction.Functioning right atrial (ra) and left ventricular (lv) leads were present in the patient as well, not targeted for extraction.A spectranetics lead locking device (lld) was inserted into the rv lead to provide traction.Multiple spectranetics devices (13f tightrail guardian motorized dilator sheath, 16f glidelight laser sheath, and 13f tightrail rotating dilator sheath (long)) were used during the procedure.While the 13f tightrail long was being used and while significant traction was being applied, the rv lead and lld separated.The physician then removed the rv lead/lld remnants through the femoral vein using a snare.The procedure was completed with no reported patient harm.This event captures the lld which separated along with the rv lead, requiring intervention.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPECTRANETICS LEAD LOCKING DEVICE
Common Device Name
STYLET, CATHETER
Manufacturer (Section D)
THE SPECTRANETICS CORPORATION
9965 federal drive
colorado springs CO 80921
Manufacturer (Section G)
SPECTRANETICS
9965 federal drive
colorado springs CO 80921
Manufacturer Contact
barbara creel
9965 federal drive
colorado springs, CO 80921
719447-2465
MDR Report Key16499811
Report Number1721279-2023-00038
Device Sequence Number15048483
Product Code DRB
UDI-Device Identifier00813132023010
UDI-Public00813132023010
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K990713
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/07/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number518-019
Device Catalogue Number518-019
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date03/07/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
MEDTRONIC 4298 LV LEAD.; MEDTRONIC 5076 RA PACING LEAD.; MEDTRONIC 6935M RV ICD LEAD.; SPECTRANETICS 16F GLIDELIGHT LASER SHEATH.; SPECTRANETICS CVX-300 EXCIMER LASER SYSTEM.; SPECTRANETICS TIGHTRAIL GUARDIAN DILATOR SHEATH.; SPECTRANETICS TIGHTRAIL ROTATING DILATOR SHEATH.
Outcome Attributed to Adverse Event Required Intervention;
Patient Age67 YR
Patient SexMale
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-