• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GYRUS ACMI, INC. VALVE IN CARTRIDGE, 5MM; SPIRATION VALVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GYRUS ACMI, INC. VALVE IN CARTRIDGE, 5MM; SPIRATION VALVE Back to Search Results
Model Number SVS-V5-00
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Pneumothorax (2012)
Date of Event 01/31/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Four spiration valves (1x svs-v7-00, 2x svs-v9-00, 1x svs-v5-00) were placed in the patient.A total of four event reports were filed separately, one report for each device.This is the second of four reports.Pneumothorax is the most common device related serious adverse event that is associated with the spiration valve system.In the emprove study, the incidence of serious pneumothorax was 14.2%, with over 75% of the serious pneumothorax events occurring within the first 3 days post-procedure.Early-onset pneumothorax in the treatment group likely resulted from lung conformation changes due to acute reduction in lung volume by valve therapy, triggering rapid expansion of the ipsilateral non-targeted lobe leading to pneumothorax.(criner gj, delage a, voelker k, et al.Improving lung function in severe heterogenous emphysema with the spiration® valve system (emprove): a multicenter, open-label, randomized control trial.Am j respir crit care med.2019;200(11):1354-1362.Doi:10.1164/rccm.201902-0383oc).However, it should be noted that pneumothorax events are also recognized as an indicator of successful target lobe occlusion and when managed according to published expert guidelines (valipour a, slebos dj, de oliveira hg, et al.Expert statement: pneumothorax associated with endoscopic valve therapy for emphysema -- potential mechanisms, treatment algorithm and case examples.Respiration 2004 2014;87(6): 513-521.Doi:10.1159/000360642), subjects with pneumothorax events experienced clinical benefits similar to that in subjects without pneumothorax events (criner gj, delage a, voelker k, et al.Improving lung function in severe heterogenous emphysema with the spiration valve system (emprove).A multicenter, open-label, randomized control trial.Am j respir crit care med.2019;200(11):1354-1362.Doi:10.1164/rccm.201902-0383oc).The reported event aligns with the experience in the emprove clinical trial and is an expected adverse event associated with the spiration valve system.
 
Event or Problem Description
Patient had four spiration valves placed on (b)(6) 23.On (b)(6) 23 patient developed a small left apical pneumothorax, which resolved without intervention within two days.Seven days after surgery patient developed a new large pneumothorax with atelectasis of the left lung and rightward mediastinal shift.Chest tube was placed and after subsequent x-ray, medical team inserted an interim left-sided pleural-based pigtail catheter in situ.Hospital stay was prolonged.On (b)(6) 23, doctor stated that patient had no air leak in the inferior chest tube, had decreased diminishing fluid, and that chest x-ray was improved.Doctor stated that the patient was overall improved.Patient was released on (b)(6) 23 with heimlich valve in place.Pneumothorax resolved, and patient returned to have the heimlich valve removed about a week later, (b)(6) 23.Spiration valves remain in place.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VALVE IN CARTRIDGE, 5MM
Common Device Name
SPIRATION VALVE
Manufacturer (Section D)
GYRUS ACMI, INC.
6675 185th ave ne
redmond WA 98052
Manufacturer (Section G)
GYRUS ACMI, INC.
6675 185th ave ne
redmond WA 98052
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key16500064
Report Number3004450998-2023-00020
Device Sequence Number12294440
Product Code NJK
UDI-Device Identifier00896506002286
UDI-Public00896506002286
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
P180007
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Study,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 03/07/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberSVS-V5-00
Device Lot NumberWS265232
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/06/2023
Initial Report FDA Received Date03/07/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
Patient Age67 YR
Patient SexMale
-
-