• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TORNIER INC PERFORM CENTERED HUMERAL HEAD COUPLER TI; PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TORNIER INC PERFORM CENTERED HUMERAL HEAD COUPLER TI; PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED Back to Search Results
Model Number DWT100
Device Problem Device Dislodged or Dislocated (2923)
Patient Problems Non-union Bone Fracture (2369); Subluxation (4525)
Event Date 10/04/2022
Event Type  Injury  
Event Description
It was reported that a patient had an anterior subluxation of the right shoulder with failure of the subscapularis repair.
 
Manufacturer Narrative
The reported event could be confirmed since x-rays were provided for evaluation that shows superior subluxation of the humerus in relation to the glenoid component.The opinion of the medical expert was requested and stated as following: ¿this case describes rotator cuff failure postoperatively after an anatomic total shoulder arthroplasty.The prosthesis is recognized as the perform stem, humeral head and pe-glenoid.There are several reasons for rotator cuff failure postoperatively.I will list them here for educational purposes as well: pre-existing rotator cuff pathology: patients with pre-existing rotator cuff tears or tendinopathy may be at higher risk of postoperative cuff failure, as the surgical intervention may cause further damage to the already compromised tissue.Failure of surgically released subscapularis tendon to heal postoperatively.Surgical technique: the surgical technique used during the procedure can also affect the risk of rotator cuff failure.Improper placement of the implant or excessive tension on the cuff during implantation can lead to cuff failure.Implant design: the design of the implant used in the procedure can also play a role in the development of rotator cuff failure.Implants that are too large or too small or those that do not match the patient's anatomy can cause mechanical stress on the cuff and lead to failure.Inadequate rehabilitation: proper rehabilitation following surgery is essential for the successful recovery of the rotator cuff.Failure to follow a prescribed rehabilitation program or returning to activities too soon can put additional stress on the cuff and lead to failure.Age: older patients may be at higher risk for rotator cuff failure due to a decreased ability of the tissue to heal and regenerate.Other medical conditions: patients with diabetes, osteoporosis, or other chronic medical conditions may be at higher risk for rotator cuff failure due to compromised healing capacity of the tissue.The x-ray shows a superior subluxation of the humerus in relation to the glenoid component, indication (also) superior rc insufficiency.If you look carefully, you can see the glenoid component as a negative contrast (dark structure).The x-ray shows correct placement and sizing of the shoulder arthroplasty components.In this case we can think of items 1 and 2 as the most probable explanation for this event.Most likely patient-related, possibly user-related.¿ a device inspection was not possible since the affected device was not returned, and only x-rays were provided for the investigation.The batch record could not be reviewed because the affected device was not returned, and the lot number was not communicated.A review of the labeling did not indicate any abnormalities.More information as well as the affected device must be available in order to determine the exact root cause of the alleged failure in order to confirm that it is patient related.If any additional information is provided, the investigation will be reassessed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERFORM CENTERED HUMERAL HEAD COUPLER TI
Type of Device
PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED
Manufacturer (Section D)
TORNIER INC
10801 nesbitt avenue s
bloomington MN 55437
Manufacturer (Section G)
TORNIER INC
10801 nesbitt avenue s
bloomington MN 55437
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key16501092
MDR Text Key310854989
Report Number0001649390-2023-00028
Device Sequence Number1
Product Code HSD
UDI-Device Identifier00846832085088
UDI-Public00846832085088
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K201315
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial
Report Date 03/08/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/08/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model NumberDWT100
Device Catalogue NumberDWT100
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/07/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age57 YR
Patient Weight86 KG
-
-