Model Number 86-4183
Medical Device Problem Code Loss of or Failure to Bond (1068)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 02/23/2023
Type of Reportable Event Serious Injury
Event or Problem Description
Revision of a pfc tkr done in 2006.Male, left side.On the joint registry form, the reason for revision was due to deep infection, tibial loosening, femur loosening.When exposure was done the surgeon noted significant poly wear as seen in the pics.All components were removed and attune revision was implanted.The patient notes were unavailable so i was unable to get the lot numbers from the implant stickers.The only lot number i was able to read on the implants was for the one stated in the impacted products.86-4183 pfc keeled tibial tray sz 5.The tibia was very loose.As you can see in the pic there is cement on the implant, so both interfaces failed.The femur failed at the bone-cement interface.
 
Additional Manufacturer Narrative
Product complaint # (b)4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint (b)(4).Investigation summary; the device associated with this report was not returned to depuy synthes for evaluation.Review of the photographic evidence was not able to confirm the complaint.No series of chronological x-ray was provided and no signs of implant loosening on the explanted device were observed.Extraction marks can be observe on the tibial tray.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot; the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records (mre) was not performed.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: the device associated with this report was not returned to depuy synthes for evaluation.Review of the photographic evidence was not able to confirm the complaint.No series of chronological x-ray was provided and no signs of implant loosening on the explanted device were observed.Extraction marks can be observe on the tibial tray.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PFC KEEL TIB TRAY CEM SZ5
Common Device Name
SIGMA KNEE PRIMARY : KNEE TIBIAL TRAY
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY IRELAND 9616671
loughbeg, ringaskiddy co.
cork
EI  
Manufacturer Contact
kate karberg
700 orthpaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key16501657
Report Number1818910-2023-05278
Device Sequence Number19671136
Product Code JWH
UDI-Device Identifier10603295216452
UDI-Public10603295216452
Combination Product (Y/N)N
Initial Reporter CountryNZ
PMA/510(K) Number
K961685
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2006
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 03/08/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Model Number86-4183
Device Catalogue Number864183
Device Lot Number2067657
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/23/2023
Supplement Date Received by Manufacturer03/10/2023
03/14/2023
05/17/2023
Initial Report FDA Received Date03/08/2023
Supplement Report FDA Received Date03/13/2023
03/17/2023
05/18/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/17/2006
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
PFC*SIGMA C/R NPOR FEM LT SZ5; SIGMA CRVD GVF INS 5 10MM; UNKNOWN BONE CEMENT
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-