MAUDE Adverse Event Report: SYNTHES GMBH MATRIXMANDIBLE 1.5MM DRILL BIT J-LATCH FOR 03.503.045/.047; DRILL, BONE, POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH MATRIXMANDIBLE 1.5MM DRILL BIT J-LATCH FOR 03.503.045/.047; DRILL, BONE, POWERED Back to Search Results
Model Number 03.503.476
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/17/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Depuy synthese is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthese has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthese or its employees that the report constitutes an admission that the device, depuy synthese, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Device report from colombia reports an event as follows: it was reported that on (b)(6) 2023, the drill bit in question fractured.There was no reported adverse patient impact.No further information is available.This report is for a matrixmandible 1.5mm drill bit j-latch for 03.503.045/.047.This is report 1 of 1 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: h6 part # 03.503.476.Synthes lot # u381642.Supplier lot # u381642.Release to warehouse date: 24 jun 2021.Supplier : (b)(4).No non conformance reports were generated during production.The photo was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device visual analysis of the photo revealed that the matrixmnd 1.5 db j-latch was found broken from the drilling shaft.The broken portion is shown on the evidence provided.¿the observed condition of the device was consistent with a random component failure that may have been caused by exposure to unintended forces.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed for the matrixmnd 1.5 db j-latch.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MATRIXMANDIBLE 1.5MM DRILL BIT J-LATCH FOR 03.503.045/.047
Common Device Name
DRILL, BONE, POWERED
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
DEPUY MOTION SARL
chemin blanc 38
le locle
SZ  
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key16504583
Report Number8030965-2023-02881
Device Sequence Number14291208
Product Code DZI
UDI-Device Identifier10887587012971
UDI-Public(01)10887587012971
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
K043310
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number03.503.476
Device Catalogue Number03.503.476
Device Lot NumberU381642
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/17/2023
Supplement Date Received by Manufacturer04/04/2023
Initial Report FDA Received Date03/08/2023
Supplement Report FDA Received Date04/20/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/24/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-