• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET 3I LOCATOR¿ ABUTMENT 4.1MM(D) X 3MM(H); DENTAL ABUTMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET 3I LOCATOR¿ ABUTMENT 4.1MM(D) X 3MM(H); DENTAL ABUTMENT Back to Search Results
Model Number LOA003
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Zimmer biomet complaint number (b)(4).Date of event unknown / not provided.
 
Event or Problem Description
It was reported that the abutment at tooth site 45 fractured at the screw of the abutment and was removed.
 
Event or Problem Description
No further event information is available at the time of this report.
 
Additional Manufacturer Narrative
Zimmer biomet complaint number (b)(4).The complaint report was received, reviewed, and evaluated by the manufacturer's complaint process.Applicable manufacturer¿s internal lot history records were reviewed to evaluate whether any internal deviations, non-conformance's, or problems occurred during the manufacture of the lot, which could cause or contribute to the reported complaint condition.A review of other manufacturers complaint files and related trending data for similar incidents was performed to evaluate whether other similar complaints have been received and if data suggests any adverse trends may have contributed to the complaint root cause.Condition of the returned complaint related product was evaluated to confirm the reported product complaint.As part of each complaint investigation, the manufacturer performs an analysis to determine the root-cause of the reported problem.Depending upon the information provided, a true root-cause may or may not be determined.In these cases, the most-likely cause of the reported issue would be determined.Other known issues that are inherent to the business operations, typically ones that are related to shipping, handling, customer service, or other, are internally tracked and continually trended for unusual increases.As a result of the above investigation, the manufacturer has concluded the following as the root-cause or most likely root-cause of the reported complaint condition: no manufacturing event was confirmed, event is tracked and trended with no further actions taken at this time.Most likely hazard experience: thread fracture during function.Most likely root cause of the failure mode / hazard: thread fracture during function typically results from insufficient tightening torque at placement, overloading of the abutment in function, or not retightening the abutment at prescribed intervals.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LOCATOR¿ ABUTMENT 4.1MM(D) X 3MM(H)
Common Device Name
DENTAL ABUTMENT
Manufacturer (Section D)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
susanne taylor
4555 riverside drive
palm beach gardens, FL 33410
5617766700
MDR Report Key16504864
Report Number0001038806-2023-00453
Device Sequence Number11588641
Product Code DZE
UDI-Device Identifier00844868009559
UDI-Public(01)00844868009559(17)240609(10)1229593
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
K012911
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date (Section B) 04/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date06/09/2024
Device Model NumberLOA003
Device Catalogue NumberLOA003
Device Lot Number1229593
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/13/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/13/2023
Supplement Date Received by Manufacturer03/22/2023
Initial Report FDA Received Date03/08/2023
Supplement Report FDA Received Date04/10/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/11/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient Age78 YR
Patient SexFemale
Patient Weight70 KG
-
-