Model Number FX385T
Medical Device Problem Code Infusion or Flow Problem (2964)
Health Effect - Clinical Code Hydrocephalus (3272)
Date of Event 02/04/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Manufacturing site evaluation: no product received to date.Should relevant additional information/investigation results become available, an additional medwatch report will be submitted.
 
Event or Problem Description
It was reported that a progav 2.0 shunt system (part # fx385t) was implanted during a procedure performed on (b)(6) 2023.According to the complainant, approximately two (2) days after being implanted, the valve setting was not able to be reprogrammed.The surgeon made several unsuccessful attempts to perform the adjustment with the sales representative present.The patient underwent a revision procedure.The complaint device has not yet been returned to the manufacturer for evaluation.No patient complications were reported as a result of the revision procedure.
 
Additional Manufacturer Narrative
Manufacturing site evaluation: information about the complaint: information about the complaint are: "malfunction" manufacturing and quality control data: due to the limited information and the missing product investigation is restricted to tracability of manufacturing and quality control data.The progav 2.0 shunt system was manufactured by a qualified employee on march 2022.Deviations during assembly did not occur.The product was finally tested as article fx385t and released for packaging and sterilization and was sterilized by miethke.The progav 2.0 has a normal pressure range of 0 to 20 cmh2o.The parameters after completion of the valve assembly were tested at a flow rate of 5 ml/h or 50 ml/h at set pressures of 0, 5, 10 and 20 cmh2o, and were found to meet specifications.The valve was pre-adjusted to 5 cmh2o in accordance with the specification.The shuntassistant 2.0 has a fixed pressure of 20 cmh2o.The parameters after completion of the valve assembly were tested at a flow rate of 5 ml/h or 50 ml/h at a fixed pressure of 20 cmh2o, and were found to meet range the tolerances.Conclusion information: an investigation was not possible because no product was sent for investigation.On the basis of the product documentation, we can exclude the presence of a defect at the time of delivery of the progav 2.0 shuntsystem, since this documentation indicates that the valve is in perfect condition.Deposits caused by natural substances in the cerebrospinal fluid, such as protein, blood or tissue particles, are among the known and unavoidable risks and side effects of hydrocephalus therapy.Small amounts of non-visible deposits/proteins might compromise the integrity of the valve.Unfortunately, due to the missing sample, we are unable to provide a meaningful analysis.Further actions: from our point of view, no further regulatory actions are required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGAV 2.0 W/SA2.0 20 & SPRUNG RESERVOIR
Common Device Name
HYDROCHEPHALUS MANAGEMENT
Manufacturer (Section D)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM  14469
Manufacturer (Section G)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM   14469
Manufacturer Contact
joerg knebel
ulanenweg 2
potsdam, 14469
GM   14469
MDR Report Key16507353
Report Number3004721439-2023-00049
Device Sequence Number19671389
Product Code JXG
UDI-Device Identifier04041906502461
UDI-Public4041906502461
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K190174
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberFX385T
Device Catalogue NumberFX385T
Device Lot Number20058121
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/06/2023
Supplement Date Received by Manufacturer03/15/2023
Initial Report FDA Received Date03/08/2023
Supplement Report FDA Received Date04/13/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/15/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-