MAUDE Adverse Event Report: SHIRAKAWA OLYMPUS CO., LTD. HIGH FLOW INSUFFLATION UNIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SHIRAKAWA OLYMPUS CO., LTD. HIGH FLOW INSUFFLATION UNIT Back to Search Results
Model Number UHI-4
Device Problem No Visual Prompts/Feedback (4021)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The device was returned to olympus for evaluation and the customer's allegation was confirmed.The device evaluation found the seven segment bar light emitting diode (led) was dead/not working.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation.
 
Event Description
The customer reported to olympus, the high flow insuffluation unit digits on volume indicator was faulty.The seven digit segment bar was not working.Insuffluation seemed unaffected.The issue was found during a therapeutic procedure.Inspection and testing of the returned device found the seven segment bar light emitting diode (led) was not working.There were no reports of patient harm.This medical device report (mdr) is being submitted to capture the reportable malfunction found during evaluation.
 
Event Description
There was no delay or postponement of the procedure.The procedure was a cholecystectomy.No error messages were displayed.The device was replaced with a loaner unit for the duration of the repair.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the definitive root cause of the faulty light-emitting diode (led) parts could not be determined.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIGH FLOW INSUFFLATION UNIT
Type of Device
HIGH FLOW INSUFFLATION UNIT
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key16507536
MDR Text Key311294384
Report Number3002808148-2023-02280
Device Sequence Number1
Product Code HIF
UDI-Device Identifier04953170324147
UDI-Public04953170324147
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K122180
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 05/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/08/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberUHI-4
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/01/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received04/06/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/03/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-