| Model Number |
FX415T |
| Medical Device Problem Code |
Infusion or Flow Problem (2964)
|
| Health Effect - Clinical Code |
Hydrocephalus (3272)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that a shuntassistant (part # fx415t) was implanted during a procedure performed on an unknown date.According to the complainant, the shunt system was believed to be operating in underdrainage.The patient underwent a revision procedure.The complaint device was returned to the manufacturer for evaluation.No patient complications were reported as a result of the revision procedure.Age: 6 years, height: 120 centimeters (cm), weight: 30 kilograms (kg), gender: male.Same event as medwatch # 3004721439-2023-00039.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation: visual inspection: during the investigation, no significant deformations or damage of the valves was determined.Permeability test: a permeability test has shown that the valve is permeable.Computer controlled test: to investigate the claim of under-drainage, the opening pressure is measured using a miethke computer controlled testing apparatus which simulates a cerebrospinal fluid flow.The valve is tested in the vertical position.The results show that the shuntassistant operates within the accepted tolerance in the horizontal position.Internal inspection: after dismantling of the valve, deposits were found in shuntassistant.Results: based on our investigation results, we can determine deposits.The deposits had no effect to the technical properties at the time of the investigation.The cause of the aforementioned functional impairment is not known to us at the time of the examination.Proteins in the cerebrospinal fluid can influence the function temporarily and are known side effects in hydrocephalus therapy.We can exclude a defect at the time of release.The valve met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.No further regulatory actions are required from our point of view.
|
| |
|
Search Alerts/Recalls
|