Investigation: visual inspection: during the investigation, no significant deformation or damage of the valve was determined.Permeability test: a permeability test has shown that the prosa is permeable.Computer controlled test: to investigate the claim of under-drainage, the opening pressure is measured using a miethke computer controlled testing apparatus which simulates a cerebrospinal fluid flow.The valve is tested in the vertical position.The results show that the valve is operating not within the accepted tolerance in the vertical position.An accelerated outflow of prosa could be determined.Adjustment test: the prosa valve was tested and is adjustable to all specified pressures.Braking force and brake function test: the brake functionality test has shown that the brake function is fully operational and the braking force is within the given tolerances.Internal inspection: after dismantling of the valve, deposits were found in prosa.Results: based on our investigation results, we can determine an accelerated outflow in the valve.The determined deposits can be named as the cause for the accelerated outflow.Proteins in the cerebrospinal fluid can influence the function temporarily and are known side effects in hydrocephalus therapy.We can exclude a defect at the time of release.The valve met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.No further regulatory actions are required from our point of view.
|