• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CONMED UTICA ELEC,BLDE,LNG,EXT,INSULATD,40; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CONMED UTICA ELEC,BLDE,LNG,EXT,INSULATD,40; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number 138110
Medical Device Problem Code Delivered as Unsterile Product (1421)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Received two 138110 in original packaging.Lot number was verified.Performed a visual inspection, there were no obvious signs of a breach in the seal on one package however, the device was encroaching into the seal on the second device.Performed a functional inspection, the devices were dye leak tested per tm-10-217 rev.F which indicated that the packaging had an insufficient heat seal on one out of the two packages.The manufacturing documents from the device history record have been reviewed with special attention to the manufacturing and inspection of the product.The product released for distribution was found to have met all specifications prior to shipment.This is the only complaint for this lot number and failure mode within the past two years.A two-year review of complaint history revealed there has been a total of 101 complaints, regarding 7543 devices, for this device family and failure mode.During this same time frame (b)(4) devices have been manufactured and shipped worldwide.(b)(4).Per the instructions for use, the user is advised the following: these devices should be inspected before and after each use.Visually examine the devices for obvious physical damage including: cracked, broken or otherwise distorted plastic parts.Damage including cuts, punctures, nicks, abrasions, unusual lumps, significant discoloration.We will continue to monitor for trends through the complaint system to assure patient safety.
 
Event or Problem Description
During incoming inspection, the distributor rejected this device,138110, for an insufficient heatseal.There was no contact with the patient as this was found during incoming inspection.This will be reported as a malfunction with potential for injury upon reoccurrence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELEC,BLDE,LNG,EXT,INSULATD,40
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
CONMED UTICA
525 french rd
utica NY 13502
Manufacturer (Section G)
CONMED UTICA
525 french rd
utica NY 13502
Manufacturer Contact
samantha dewberry
525 french road
utica, NY 13502
3152230184
MDR Report Key16511592
Report Number1320894-2023-00051
Device Sequence Number14915165
Product Code GEI
UDI-Device Identifier30653405003452
UDI-Public(01)30653405003452(17)260718(10)202107191
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K943542
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date (Section B) 03/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Expiration Date07/18/2026
Device Model Number138110
Device Catalogue Number138110
Device Lot Number202107191
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/27/2023
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/16/2023
Initial Report FDA Received Date03/09/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/19/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
Patient EthnicityNon Hispanic
-
-