| Model Number |
C-LINE CARPAL TUNNEL SCOPE,30°,2.9X158MM |
| Medical Device Problem Codes |
No Display/Image (1183); Defective Device (2588)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
02/17/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2023 it was reported by a sales representative via sems that an ar-3350-2930 scope view is obstructed.This was discovered during an endoscopic carpal tunnel release and was resolved by using an open procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: g3, h3, h6.Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause of this failure is attributed to wear and tear damage incurred over repeated usage.Manufactured date 2018.
|
| |
|