• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION RX CYTOLOGY BRUSH; ENDOSCOPIC CYTOLOGY BRUSH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION RX CYTOLOGY BRUSH; ENDOSCOPIC CYTOLOGY BRUSH Back to Search Results
Model Number M00545000
Medical Device Problem Codes Detachment of Device or Device Component (2907); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/10/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported to boston scientific corporation that an rx cytology brush was used during an endoscopic retrograde cholangiopancreatography procedure performed on (b)(6) 2023.During the procedure, the tip of the brush was damaged.The procedure was completed with a similar rx cytology brush.There were no patient complications reported as a result of this event.
 
Additional Manufacturer Narrative
Imdrf device code a0501 captures the reportable event of brush detached.
 
Additional Manufacturer Narrative
Block h6: imdrf device code a0501 captures the reportable event of brush detached.Block h10: investigation results an rx cytology brush was received for analysis, and a visual inspection of the returned device revealed the handle and working length were kinked.No other problems were noted the reported event of brush detached was not confirmed because the results of the analysis performed showed that the brush was in good condition.Most likely, procedural factors such as the handling of the device and the technique used by the physician during initial use, set-up, or shortly thereafter could have affected its condition and caused the handle and the working length to kink.Based on the information available and the returned device analysis, the most probable root cause for the reported complaint is no problem detected and the most probable root cause for the failures found during analysis (working length - bent/kinked) and (handle - bent/kinked) is adverse event related to procedure.
 
Event or Problem Description
It was reported to boston scientific corporation that an rx cytology brush was used during an endoscopic retrograde cholangiopancreatography procedure performed on (b)(6) 2023.During the procedure, the tip of the brush was damaged.The procedure was completed with a similar rx cytology brush.There were no patient complications reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RX CYTOLOGY BRUSH
Common Device Name
ENDOSCOPIC CYTOLOGY BRUSH
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
780 brookside drive
spencer IN 47460
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key16512172
Report Number3005099803-2023-01021
Device Sequence Number20428596
Product Code FDX
UDI-Device Identifier08714729268628
UDI-Public08714729268628
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K930348
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/23/2024
Device Model NumberM00545000
Device Catalogue Number4500
Device Lot Number0029459808
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/13/2023
Supplement Date Received by Manufacturer04/18/2023
Initial Report FDA Received Date03/09/2023
Supplement Report FDA Received Date05/12/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/24/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age75 YR
Patient SexMale
Patient Weight62 KG
-
-