• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDOS INTERNATIONAL SARL 4 WAY FOOT PEDAL FOR FMS DUO+; FOOT-SWITCH, ELECTRICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SARL 4 WAY FOOT PEDAL FOR FMS DUO+; FOOT-SWITCH, ELECTRICAL Back to Search Results
Model Number 284176
Medical Device Problem Code Output Problem (3005)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Additional narrative: udi: (b)(4).Reporter is a j&j employee.Investigation summary
=
> the complaint device was received at the service center and evaluated.During the service evaluation the following defects were identified: ¿ functional : intermittent operation.Per service reports, this complaint can be confirmed.The device was scrapped as system is non-repairable.As part of depuy mitek¿s quality process all devices are manufactured, inspected, and released to approved specifications.The faulty parts was identified as the root cause for the device failure during the service evaluation.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post market surveillance.At this point in time, no corrective action is required, and no further action is warranted.Depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Device history review
=
> manufacturing record evaluation is not required as the reported event is not associated with the manufacturing process and/or the potential cause of the defect cannot be associated to manufacturing.
 
Event or Problem Description
It was reported by the affiliate in italy that preoperatively to an unknown procedure on an unknown date, it was observed that the 4 way foot pedal for fms duo+ device was broken.During in-house engineering evaluation, it was determined that the device had an intermittent operation.It was unknown if and how the procedure was completed.There were no adverse patient consequences nor surgical delay reported.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
4 WAY FOOT PEDAL FOR FMS DUO+
Common Device Name
FOOT-SWITCH, ELECTRICAL
Manufacturer (Section D)
MEDOS INTERNATIONAL SARL
chemin blanc 38
le locle CH-24 00
SZ  CH-2400
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 02767
3035526892
MDR Report Key16513197
Report Number1221934-2023-01121
Device Sequence Number6296764
Product Code HRX
UDI-Device Identifier10886705017256
UDI-Public10886705017256
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
K954465
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number284176
Device Catalogue Number284176
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/20/2023
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/20/2023
Initial Report FDA Received Date03/09/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/12/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-