• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NUVASIVE SPECIALIZED ORTHOPEDICS, INC. MAGEC GENERATION 2B SPINAL BRACING AND DISTRACTION SYSTEM; GROWING ROD SYSTEM - MAGNETIC ACTUATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NUVASIVE SPECIALIZED ORTHOPEDICS, INC. MAGEC GENERATION 2B SPINAL BRACING AND DISTRACTION SYSTEM; GROWING ROD SYSTEM - MAGNETIC ACTUATION Back to Search Results
Model Number MC2-5070R
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Metal Related Pathology (4530)
Date of Event 03/02/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device has not been returned for evaluation.The root cause is unable to be determined at this time.A review of the device history records (dhr) confirmed the device met all quality inspections and specifications prior to release.If any additional information is provided, a supplemental report will be submitted.
 
Event or Problem Description
Information was received that upon removal there was metallosis present.
 
Event or Problem Description
Information was received that during removal for definitive fusion after treatment goals were achieved, it was noted there was "metallosis" present.
 
Additional Manufacturer Narrative
The device has not been returned to nuvasive for evaluation.Investigation has been conducted using provided x-ray images.It was noted that the rod had been partially distracted.There were no obvious device-related abnormalities noted.Based on the provided information, no objective evidence has been found to indicate the presence of metallosis.The presence of metallosis cannot be verified using the radiographic pictures or x-ray images that have been provided.The reported event cannot be confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAGEC GENERATION 2B SPINAL BRACING AND DISTRACTION SYSTEM
Common Device Name
GROWING ROD SYSTEM - MAGNETIC ACTUATION
Manufacturer (Section D)
NUVASIVE SPECIALIZED ORTHOPEDICS, INC.
101 enterprise dr. ste 100
aliso viejo CA 92656
Manufacturer (Section G)
NUVASIVE SPECIALIZED ORTHOPEDICS, INC.
101 enterprise dr. ste 100
aliso viejo CA 92656
Manufacturer Contact
geoff gannon
101 enterprise dr. ste 100
aliso viejo, CA 92656
8583448112
MDR Report Key16513658
Report Number3006179046-2023-00155
Device Sequence Number6411149
Product Code PGN
UDI-Device Identifier00887517834010
UDI-Public887517834010
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K171791
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/31/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberMC2-5070R
Device Lot Number7090903AAA
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/02/2023
Supplement Date Received by Manufacturer03/30/2023
Initial Report FDA Received Date03/09/2023
Supplement Report FDA Received Date03/31/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/09/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Other;
Patient Age11 YR
Patient SexFemale
Patient Weight43 KG
-
-