PHILIPS IMAGE GUIDED THERAPY CORPORATION TIGHTRAIL GUARDIAN MOTORIZED DILATOR SHEATH; DILATOR, VESSEL, FOR PERCUTANEOUS CATHETERIZATION
|
Back to Search Results |
|
| Model Number |
575-011 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/21/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Patient''s weight unk.The evaluation and investigation is ongoing, therefore conclusion not yet available.A supplemental mdr will be submitted upon completion.Submission of this report does not, in itself, represent a conclusion by the manufacturer and/or authorized representative or the national competent authority that the content of this report is complete or accurate, that the medical device(s) listed failed in any manner and/or that the medical device(s) caused or contributed to an alleged death or deterioration in the state of the health of any person.
|
| |
|
Event or Problem Description
|
|
A lead extraction procedure commenced to remove a right atrial (ra) and a right ventricular (rv) lead due to malfunction.A left ventricular (lv) lead was present in the patient as well but was not targeted for extraction.Spectranetics lead locking devices (llds) were inserted into both leads to provide traction; however neither lld reached the distal tip of the leads, but were adequate to provide traction.The physician began by using a spectranetics 11f tightrail guardian motorized dilator sheath and while attempting to extract the ra lead, the device inadvertently cut the ra lead, and the lld pulled out of the lead.Attempt was then made to extract the rv lead using a 13f tightrail guardian when the device inadvertently cut the rv lead, and the lld pulled out of the lead (mdr #1721279-2023-00048).The physician then moved to the femoral vein where an abbott agilis introducer and a bioptome catheter were used to successfully retrieve both leads.The procedure was completed with no reported patient harm.This event captures the 11f tightrail guardian which inadvertently cut the ra lead, requiring intervention.There was no alleged malfunction of any spectranetics devices in use during the procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
B5/h2): correction: this event is no longer reportable per physician assessment completed on 13apr2023.D1): correction: removed "spectranetics" from brand name.Manufacturer name changed from "the spectranetics corporation" to "philips image guided therapy corporation".G3): review of event was completed 10apr2023.H3): the tightrail guardian and its outer sheath were returned for evaluation.The outer sheath was removed from the tightrail guardian, and appeared to be cut, approximately 14.5 inches from the beveled end.Visualizing the device, biologics were present along the length of the device, distal tip, and distal band, with a slight downward bend in the shaft, approximately 8-10 inches from the distal tip.A cracking sound was heard at the distal tip when the device initialized, indicating biologics breaking loose.Fluid ingress into the handle cavity was noted near the des flex mode switches, but there was no electrical interaction with the switches, and this minimal leak was observed to be distal of the des flex switch board.There was no interaction between biologics and any electrical components.There was no major damage to the device, nor any design, process, or use related failure identified.H6): based on the physician''s assessment, this is not considered an intervention to preclude permanent impairment of a body function or permanent damage to a body structure.Certain h6 codes submitted on the initial mdr are no longer applicable and updated with: heic code 2199 replaces 4641 and 4621, investigation findings code 3221 replaces 3233, and investigation code 67 replaces 11.Submission of this report does not, in itself, represent a conclusion by the manufacturer and/or authorized representative or the national competent authority that the content of this report is complete or accurate, that the medical device(s) listed failed in any manner and/or that the medical device(s) caused or contributed to an alleged death or deterioration in the state of the health of any person.
|
| |
|
Event or Problem Description
|
|
This event is no longer reportable for the tightrail guardian device causing or contributing to an adverse event requiring intervention.After further review and physician assessment, it was determined that snaring is a standard adjunctive technique in lead extraction procedures, not considered additional intervention.
|
| |
|