| Model Number |
M00561191 |
| Medical Device Problem Code |
Device Contamination with Chemical or Other Material (2944)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/01/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(date of event): approximated based on the date the manufacturer became aware of the event.Imdrf device code a180104 captures the reportable event of foreign material found in device.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that a 10mm captivator ii round stiff snare was examined during a routine product check performed on an unknown date.During a routine product check, the nurse clinician found a particle in the unopened packaging.No further information has been obtained despite good faith efforts.There was no patient involvement with this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Block b3 (date of event): approximated based on the date the manufacturer became aware of the event.Block h6: imdrf device code a180104 captures the reportable event of foreign material found in device.Block h10: investigation results: one captivator snare was received for analysis.Visual analysis of the device noted that a foreign cardboard material was found inside the sealed pouch.No other device problems were noted.The reported event of foreign material present in the device was confirmed since it was observed that a foreign cardboard material was found inside the sealed pouch.The product record review confirmed this was not a new failure type, and the risk was anticipated.There was no evidence of a manufacturing problem, design, or user problem which could have caused the complaint.Based on the analysis of the returned device and the information available, the most probable cause is manufacturing deficiency.This supplemental report was not able to be submitted on time by boston scientific because of delayed acknowledgments from fda for the initial report.An fda system issue resulted in delayed acknowledgments.This report will not be considered late because it resulted from an fda system issue.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that a 10mm captivator ii round stiff snare was examined during a routine product check performed on an unknown date.During a routine product check, the nurse clinician found a particle in the unopened packaging.No further information has been obtained despite good faith efforts.There was no patient involvement with this event.
|
| |
|
Search Alerts/Recalls
|