Model Number HYF-V
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
The customer reported to olympus, that during an unspecified procedure the visera hysterovideoscope insertion section was damaged.There were no reports of patient harm associated with this event.During evaluation of the returned device, the evaluation found that the forcep stopper cap was scraping.This mdr is being submitted to capture the reportable malfunction found during the device evaluation.
 
Additional Manufacturer Narrative
The device was returned to olympus for evaluation.The customer's complaint was not confirmed.In addition to the findings reported in description of event or problem, the following non-reportable malfunctions were identified: water tightness lost, due to pinhole on connecting tube; adhesive on rubber chipped; angulation lever does not move smoothly, due to damage; and forceps cannot be smoothly inserted, due to damage on channel tube.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, a root cause could not be determined.However, it is likely that the cause of the phenomenon is mishandling of the user or that wear of ordinary use led to breakage of the device.The following is included in the device ifu: "do not bend, hit, drag, twist or drop the tip of the endoscope, insertion section, bending operation section, video cable and the light guide connector with strong force for preventing failure, breakage of the tip of the endoscope, flexible part, bending part, operation part, universal cord, and device and drop of parts." olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VISERA HYSTEROVIDEOSCOPE
Common Device Name
HYSTEROVIDEOSCOPE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key16518920
Report Number3002808148-2023-02329
Device Sequence Number3290451
Product Code HIH
UDI-Device Identifier04953170411564
UDI-Public04953170411564
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K221557
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 04/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberHYF-V
Device Catalogue NumberN6007430
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/07/2023
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/07/2023
Supplement Date Received by Manufacturer03/21/2023
Initial Report FDA Received Date03/09/2023
Supplement Report FDA Received Date04/11/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/05/2007
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-