| Model Number |
WA2UR12A |
| Medical Device Problem Code |
Fitting Problem (2183)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/06/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus and the customer's allegation was not confirmed.There was fiber breakage and there was a dent on the distal edge.Additional information has been requested regarding this event.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation or when additional information becomes available.
|
| |
|
Event or Problem Description
|
|
The customer¿s olympus representative reported to olympus technical assistance center (tac), the luer lock would not fit on the ureteroscope.The luer lock was attempted on other scopes and it worked.Multiple luer locks were attempted on the subject device and none of luer locks worked.The olympus representative reported that there might be damage to the connector for the luer lock.There were no reports of patient harm associated with this event.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, it is likely that the reportable phenomenon can be attributable to the use of excessive force by the customer.Olympus will continue to monitor field performance for this device.
|
| |
|