| Model Number |
SN6AT4 |
| Medical Device Problem Codes |
Failure to Advance (2524); Defective Device (2588)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
02/03/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A sample device was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.There have been no other complaints reported in the lot number.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
A nurse reported during the intraocular lens implantation the lens was rigid and stiff and did not slide forward in the cartridge.Additional information has been requested and received stating the surgery was completed with no issues.
|
| |
|
Additional Manufacturer Narrative
|
|
The account indicated the use of a qualified associated cartridge.A non-qualified viscoelastic was indicated.The root cause for the reported complaint may be related to a failure to follow the instructions for use (ifu).The viscoelastic indicated is not qualified for the lens/cartridge combination used.The ifu instructs that an company qualified delivery system and viscoelastic combination should be used.The use of an unqualified combination may cause damage to the lens and potential complications during the implantation process.The manufacturer internal reference number is: (b)(4).
|
| |
|