Model Number WA22606S
Medical Device Problem Codes Break (1069); Mechanical Problem (1384)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/18/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The subject device was not returned for evaluation.The cause of the issue is unknown at this time.In addition multiple follow ups were made to gather additional information, however, were unsuccessful as no response is received.Supplemental report(s) will be submitted should any relevant new information is available and or received.Investigation is ongoing.This report will be supplemented accordingly following investigation.
 
Event or Problem Description
Olympus received a medsun (b)(4) report with an event/ problem reported of "while resecting prostate during a transurethral resection prostate surgery the loop electrode got caught while in the patient on a urolift clip and the loop broke into 2 pieces both pieces.No harm to patient.The bipolar transurethral resection of prostate was completed.No harm was reported.No patient harm, no user injury reported.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, it is likely that the event occurred due to excessive force by the customer.Olympus will continue to monitor field performance for this device.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on customer response to follow up.Communication with the customer the following information conveyed: no error messages observed during the event.None reported.Procedure was a bipolar transurethral resection of prostate and was completed.No harm to the patient.Customer stated they do not have the device.No further information was provided.Investigation is ongoing.This report will be supplemented accordingly following investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HF-RESECTION ELECTRODE "PLASMALOOP", LOOP, MEDIUM, 24 FR., 12°-30°, PK TURIS
Common Device Name
HF-RESECTION ELECTRODE "PLASMALOOP
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key16519965
Report Number9610773-2023-00714
Device Sequence Number19456149
Product Code FAS
UDI-Device Identifier14042761085158
UDI-Public14042761085158
Combination Product (Y/N)N
Initial Reporter StateRI
Initial Reporter CountryUS
PMA/510(K) Number
K171965
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility
Initial Reporter Occupation Risk Manager
Type of Report Initial,Followup,Followup
Report Date (Section B) 06/30/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/07/2025
Device Model NumberWA22606S
Device Catalogue NumberWA22606S
Device Lot Number1000102362
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/10/2023
Supplement Date Received by Manufacturer03/15/2023
06/07/2023
Initial Report FDA Received Date03/10/2023
Supplement Report FDA Received Date04/11/2023
06/30/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/22/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UROLIFT CLIP
Patient Age63 YR
Patient SexFemale
-
-