| Model Number |
74401 |
| Medical Device Problem Codes |
Loose or Intermittent Connection (1371); Appropriate Device Problem Term/Code Not Available (3191)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/17/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer observed a leak while injecting cells into a cryomacs bag 250.40 ml of the cell product were lost.The patient was treated with backup material.The investigation based on descriptions of the customer showed that the white cap of the injection port of the connector set fell off.This was caused by a loose luer connection of the white cap of the injection port at the connector set.Possible causes are that the luer connection was not fastened enough during production of the connector set or got loose during production of the cryomacs freezing bag itself.Also transport movements can cause loosening of the connection.It can also not be excluded that it was loosened during use at the customer's site.For the cryomacs freezing bag 250 batch 7211100355, no deviation concerning the manufacturing process occurred.No anomalies occurred during the process.The incident rate for this issue is below 0.01%.
|
| |
|
Additional Manufacturer Narrative
|
|
3191 - appropriate term/code not available: opening causing substantial loss of material during filling.
|
| |
|