MAUDE Adverse Event Report: INSULET CORPORATION OMNIPOD DASH INSULIN MANAGEMENT SYSTEM; PUMP, INFUSION, INSULIN

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INSULET CORPORATION OMNIPOD DASH INSULIN MANAGEMENT SYSTEM; PUMP, INFUSION, INSULIN Back to Search Results
Model Number 18320
Device Problem Insufficient Information (3190)
Patient Problems Bacterial Infection (1735); Skin Infection (4544)
Event Date 02/20/2023
Event Type  Injury  
Manufacturer Narrative
According to the complainant the device will not be returned for investigation.We are unable to determine if any product condition could have contributed to customer's infusion site infection.Lot release records were reviewed and the product lot met all acceptance criteria.The omnipod is iso10993 compliant, undergoing cytotoxicity, sensitization, genotoxicity, hemolysis, irritation or intracutaneous reactivity, systemic toxicity, and implantation effects testing.The omnipod is sterilized with 100% ethylene oxide gas with a sterility assurance level of 10-6  per iso11135  and eo residual levels in compliance with iso10993.Each lot  is confirmed to meet requirements for non-pyrogenicity per iso10993 and sterility per iso11135 prior to release.
 
Event Description
It was reported that the patient went to the hospital and was diagnosed with a skin infection.For treatment, the patient was prescribed prescribed augmentin at 875 mg/125 mg tablets to take every 12 hours for 7 days and ibuprofen at 325 mg tablets every 4 hours.The pod was worn longer than 48 hours on the abdomen.The patient was discharged after 7 days.The pod was discarded.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OMNIPOD DASH INSULIN MANAGEMENT SYSTEM
Type of Device
PUMP, INFUSION, INSULIN
Manufacturer (Section D)
INSULET CORPORATION
100 nagog park
acton MA 01720
Manufacturer (Section G)
INSULET CORPORATION
100 nagog park
acton MA 01720
Manufacturer Contact
michael spears
100 nagog park
acton, MA 01720
9786007000
MDR Report Key16521882
MDR Text Key311065432
Report Number3004464228-2023-06756
Device Sequence Number1
Product Code LZG
UDI-Device Identifier20385082000020
UDI-Public(01)20385082000020
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K211575
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Type of Report Initial
Report Date 03/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/10/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model Number18320
Device Catalogue NumberBLE-I1-529
Was Device Available for Evaluation? No
Date Manufacturer Received03/09/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Hospitalization;
Patient Age60 YR
Patient SexFemale
-
-