• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS ELECSYS FERRITIN; FERRITIN IMMUNOLOGICAL TEST SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS ELECSYS FERRITIN; FERRITIN IMMUNOLOGICAL TEST SYSTEM Back to Search Results
Catalog Number 07027273190
Medical Device Problem Codes Low Test Results (2458); Non Reproducible Results (4029)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/08/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The qcs were in range.The level 1 calibration and level 2 calibration results were slightly below the approximate expected range.The investigation is ongoing.
 
Event or Problem Description
The initial reporter received a questionable elecsys ferritin (ferritin) result from one patient sample tested on the cobas e 801 analytical unit with serial number 20x6-04.The initial result was reported to the medical personnel and was rejected as it did not fit the medical state and the previous results of the patient.The initial result of the undiluted sample was 64.5 ng/ml.The first repeat result of the undiluted sample was 2000 ng/ml with a data flag.The second repeat result using a diluted sample was 4370 ng/ml.This repeat result was deemed correct as it is in line with the previous results and the clinical picture.
 
Additional Manufacturer Narrative
The most observed alarm in the alarm trace was "reagent probe liquid-level detection (lld)".This suggests a pre-analytic reagent handling issue.A general reagent issue was not identified.The investigation did not identify a product problem.The cause of the event could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECSYS FERRITIN
Common Device Name
FERRITIN IMMUNOLOGICAL TEST SYSTEM
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
na
indianapolis, IN 46250
3174767531
MDR Report Key16522830
Report Number1823260-2023-00761
Device Sequence Number12376598
Product Code JMG
UDI-Device Identifier04015630939688
UDI-Public04015630939688
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K971833
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2023
Device Catalogue Number07027273190
Device Lot Number65213901
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/10/2023
Supplement Date Received by Manufacturer03/20/2023
Initial Report FDA Received Date03/10/2023
Supplement Report FDA Received Date04/12/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-