| Model Number |
CREP G2 |
| Medical Device Problem Code |
Non Reproducible Results (4029)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/12/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Precision testing was performed with acceptable results.The field service engineer performed sample probe wash and air purge.The investigation is ongoing.(b)(4).
|
| |
|
Event or Problem Description
|
|
The initial reporter stated they received discrepant results for one patient sample tested with the crep2 (creatinine plus ver.2) assay on a cobas 6000 c 501 module, serial number (b)(4).The sample initially resulted in an invalid crep2 result, accompanied by a data flag.The sample was re-centrifuged and then repeated, resulting in a crep2 value of 0.92 mg/dl.This result was reported outside of the laboratory and questioned by the physician.The sample was repeated a second time, resulting in a crep2 value of 1.58 mg/dl.This result fits the clinical picture of the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the initial data flag indicates that a sample clot was detected.The investigation could not identify a product problem.The cause of the event could not be determined.The issue is consistent with a sample quality-related issue.
|
| |
|