• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE AIM HD LAPAROSCOPE, AUTOCLAVABLE, 5.4MM X 30°, 30CM; CONFOCAL OPTICAL IMAGING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE AIM HD LAPAROSCOPE, AUTOCLAVABLE, 5.4MM X 30°, 30CM; CONFOCAL OPTICAL IMAGING Back to Search Results
Model Number 0502537030
Medical Device Problem Code Optical Problem (3001)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/14/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that there was blurry image during procedure.
 
Additional Manufacturer Narrative
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
 
Event or Problem Description
It was reported that there was blurry image during procedure.
 
Additional Manufacturer Narrative
This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: lens on the scope is foggy.Probable root cause: incorrectly assembled optical train.Damage to optical train.End of life wear-out.Shipping damage.Use error.The reported failure mode will be monitored for future reoccurrence.Manufacture date is not known.H3 other text : 81.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AIM HD LAPAROSCOPE, AUTOCLAVABLE, 5.4MM X 30°, 30CM
Common Device Name
CONFOCAL OPTICAL IMAGING
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
chester rebugio
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key16524109
Report Number0002936485-2023-00206
Device Sequence Number6560721
Product Code OWN
UDI-Device Identifier07613327059373
UDI-Public07613327059373
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K142310
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number0502537030
Device Catalogue Number0502537030
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/14/2023
Supplement Date Received by Manufacturer02/14/2023
Initial Report FDA Received Date03/10/2023
Supplement Report FDA Received Date08/24/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-