| Model Number |
86700 |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/14/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that a visions pv.018 catheter was used in a fistulagram procedure.When the catheter was removed from the patient, a missing distal tip was noted.Fluoroscopy identified the distal tip was still over the wire inside the patient¿s mid-upper arm level.The wire was gently retracted with the distal tip into the sheath.Compression was applied for hemostasis on the fistula and an incision was made at the level of the sheath.The sheath was removed and the distal tip was fished out using surgical instruments.The incision was closed after all materials were removed.Fluoroscopy confirmed no piece was retained inside the patient.The patient was discharged as expected with no injury reported.This adverse event and product problem is being submitted due to the tip separation requiring surgical intervention.
|
| |
|
Additional Manufacturer Narrative
|
|
This complaint was reviewed and investigated according to the manufacturer¿s policy.The visions pv.018 catheter was discarded by the facility, thus no returned product investigation was performed.Submission of this report does not, in itself, represent a conclusion by the manufacturer and/or authorized representative or the national competent authority that the content of this report is complete or accurate, that the medical device(s) listed failed in any manner and/or that the medical device(s) caused or contributed to an alleged death or deterioration in the state of the health of any person.
|
| |
|
Search Alerts/Recalls
|