| Model Number |
WA96200A |
| Medical Device Problem Codes |
Break (1069); Poor Quality Image (1408)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/30/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for evaluation.The evaluation confirmed the user report and was attributed to broken optical lenses.The device was found to have a bent outer tube, minor dents on edge, and debris under the eyepiece window.This report will be supplemented if additional information is received.
|
| |
|
Event or Problem Description
|
|
It was reported that the device has broken components and the screen on the tip of the scope is cracked.The user cannot see a clear image.No patient injury reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation and correction to h3 (the device was returned at the time of the initial report; ¿not returned to manufacturer¿ was inadvertently checked on the initial report).A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, it is likely that the event occurred due to excessive use of the device.Olympus will continue to monitor field performance for this device.
|
| |
|