GYRUS ACMI, INC. SOF-CURL URETERAL STENT, 7.0FR X 26CM; STENT, URETERAL Back to Search Results
Model Number SSC7026
Medical Device Problem Codes Corroded (1131); Degraded (1153); Peeled/Delaminated (1454); Device Contamination with Body Fluid (2317); Sharp Edges (4013)
Health Effect - Clinical Code Urinary Tract Infection (2120)
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported to olympus that the sof-curl 7.0x24, lum ss.038x3 placed during a cystoscopy was visibly defective.Initial information stated that the device had caused some urinary tract infections (uti) and impacted the patient¿s quality of life.Upon follow up with the reporter, the information was contradictory.The reporter communicated that the patient had a complex history that included neurogenic bladder and bowel due to an automobile accident in (b)(6) 2021.Due to this, the patient lost bladder control and required self catheterization.Shortly after the accident, a non-olympus device was placed to help with urinary problems.It was stated that this procedure was ¿botched¿ and the patient had consistent issues with utis.That device was taken out on an unspecified date.The olympus device was then placed for 10 days on (b)(6) 2022 for ¿peephole problems.¿ upon removal of the device, it was noted that the end of the stent was ¿corroded.¿ the ¿plastic¿ was worn down, the end was pointy, and ¿paint¿ had peeled off.Additionally, it was covered with white ¿goo.¿ the device will not be returned as the patient kept it for litigation purposes against the implanting physician.Additional information has been requested.Olympus is reporting conservatively due to insufficient and conflicting information of the relationship of the device to the uti.
 
Additional Manufacturer Narrative
The investigation is ongoing and follow up with the user facility is currently being performed.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
 
Additional Manufacturer Narrative
This supplemental report is being submitted to provide additional information obtained from the customer regarding the reported event.
 
Event or Problem Description
Additional information regarding the procedure indicated that the following procedures were performed: cystoscopy, left retrograde ureteropyelogram, left ureteroscopy, balloon dilation and insertion of a left ureteral stent.A questionable obstruction was noted pre-procedure.During the procedure, the patient was given intravenous antibiotics.The prostate was noted to have obstructing medial lobes in the prostate.No foreign bodies were noted in the bladder.During the retrograde ureteropyelogram, the patient was noted to have a severely tortuous dilated ureter to the level of the distal ureter and bladder.A wire was inserted to advance into the left renal collecting system.The cystoscope was removed and a semi rigid ureteroscope was inserted and advanced to the distal ureter.Some mucusal edema was noted but no foreign bodies or stones were noted.The scope was advanced to midureter, and the ureter was found to be dilated.Urine was collected and sent for culture.As the area was edematous, balloon dilatation was done of that area.The ureteral stent was then placed with the proximal curl he left renal pelvis and the distal curl in the bladder.The patient was taken to the recovery room in stable condition.A renal ultrasound post procedure revealed persistent hydronephrosis even with the sof-curl.The documentation also noted that the sof-curl was only in for 7 days, not 10 as previously reported.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.The device history record was unable to be reviewed for this device since the serial/lot number was not provided.However, olympus only releases products to market that meet all manufacturing specifications and final product release criteria.Based on the results of the investigation, after review of the photos provided, it was observed that the surface of the device was worn and had no tubular shape.Instead, the tube at this section had similar cuts to the ones made with a knife.Based on the review of the photos and the report of the nonconformance occurring 10 days after implantation of the device, it was determined that the reported issue could not be contributed to the manufacturing process.This supplemental report includes a correction to e1.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SOF-CURL URETERAL STENT, 7.0FR X 26CM
Common Device Name
STENT, URETERAL
Manufacturer (Section D)
GYRUS ACMI, INC.
9600 louisiana avenue north
brooklyn park MN 55445
Manufacturer (Section G)
GYRUS ACMI, INC.
9600 louisiana avenue north
brooklyn park MN 55445
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key16527185
Report Number3011050570-2023-00018
Device Sequence Number11424334
Product Code FAD
UDI-Device Identifier00821925031753
UDI-Public00821925031753
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K930733
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberSSC7026
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/10/2023
Supplement Date Received by Manufacturer03/13/2023
07/18/2023
Initial Report FDA Received Date03/10/2023
Supplement Report FDA Received Date04/12/2023
08/03/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age53 YR
Patient SexMale
-
-