SMITH & NEPHEW, INC. MIS 3.2MM X 45MM RIMMED PIN STERILE; PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Catalog Number 74016465
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/13/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event or Problem Description
It was reported that during a tka surgery, the mis 3.2mm x 45mm rimmed pin sterile device broke in situ, causing the surgeon to fish out the remaining pin, out from the bone.The procedure was resumed, after a non-significant delay, with an s+n back-up device.No further complications were reported.
 
Additional Manufacturer Narrative
H3, h6: the device was not returned for evaluation.However, the photograph was reviewed, and revealed that the device is fractured.The clinical/medical investigation concluded that, the provided photo was reviewed and confirms the breakage of the pin.However, it does not provide any clinical insight into the reported breakage.As of the date of this medical investigation, the requested clinical documentation has not been provided for evaluation.Therefore, there were no clinical factors found which would have contributed to the reported event.Per complaint details, the remaining piece was ¿fished out the bone¿ by the surgeon.The procedure was completed after a non-significant delay using a smith and nephew back-up device in the original bone hole.No further complications were reported.Therefore, no further clinical/medical assessment is warranted at this time.Device batch number was not provided, thus, an evaluation of the manufacturing records could not be performed.A review of complaint history of the previous 12 months revealed similar events for the listed device, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the instructions for use documents for speed pins revealed that the speed pin should be aligned with the orientation of the instrument fixation hole and should touch the bone before drilling to avoid pin binding in the block.Also, using a mallet with speed pins must be avoided, as speed pins can bend and bind in blocks.The review of speed pins also revealed that single use devices should not be reused due to risks of breakage, this has been identified as a caution.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any specifications at the time of manufacture.This is a single use device, surgical technique or damage from misuse are likely potential factors that could contribute to the reported event.Based on this investigation, the need for corrective action is not indicated.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MIS 3.2MM X 45MM RIMMED PIN STERILE
Common Device Name
PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key16527339
Report Number1020279-2023-00492
Device Sequence Number16809736
Product Code HSX
UDI-Device Identifier00885556691984
UDI-Public885556691984
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
K113038
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/30/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number74016465
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/13/2023
Supplement Date Received by Manufacturer03/29/2023
Initial Report FDA Received Date03/13/2023
Supplement Report FDA Received Date03/30/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-