Model Number 1883070HS
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/13/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product analysis found, visually, the spiral wrap was broken 4.98 inches from the distal end of the inner hub when returned.The broken portion of the tip was 0.87 inches in length and there were striations on the outside diameter of the area between the tip and the spiral wrap.In addition, there were striations on the proximal outside diameter of the inner shaft.There was no damage in the inner or outer hubs.Functional testing could not be performed due to the broken state of the device.In the returned condition, there was an out of specification condition that was related to the complaint (due to physical damage).Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that, when surgeon used the microdebrider bur and while drilling halfway, the bur broke within the patient sinus.He was able to retrieve all the broken pieces from the sinus.Prior to the fault, he did not encounter any unusual encounters (no sound, no wobble of the burs, no heating up of handpiece).The patient had latent tuberculosis.An intervention was performed, the surgery uses an endoscope.Surgeon was able to see and confirm what was taken out from the sinuses.The bur was replaced with another similar product and the procedure was completed.There was no patient impact.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XPS® BUR - HIGH SPEED RAD®
Common Device Name
BUR, EAR, NOSE AND THROAT
Manufacturer (Section D)
COVIDIEN LLC
15 hampshire street
mansfield 02048
Manufacturer (Section G)
COVIDIEN LLC
15 hampshire street
mansfield 02048
Manufacturer Contact
glen belmer
6743 southpoint drive north
jacksonville, FL 32216
6122713209
MDR Report Key16527883
Report Number9612501-2023-00420
Device Sequence Number2645858
Product Code EQJ
UDI-Device Identifier00681490464604
UDI-Public00681490464604
Combination Product (Y/N)N
Initial Reporter CountrySN
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1883070HS
Device Catalogue Number1883070HS
Device Lot Number0219018560
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/27/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/13/2023
Initial Report FDA Received Date03/13/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/08/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1