| Model Number |
FINETUNER ECHO [FINETUNER 2] |
| Medical Device Problem Code |
No Device Output (1435)
|
| Health Effect - Clinical Code |
Burning Sensation (2146)
|
| Date of Event |
03/05/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2023 the external device started to smell burnt and the user's mother got a small electric shock.There was no serious injury sustained from the brief shock.The mother of the user returned the device to the subsidiary.
|
| |
|
Event or Problem Description
|
|
On 5th march 2023 the external device started to smell burnt and the user's mother got a small electric shock.There was no serious injury sustained from the brief shock.
|
| |
|
Additional Manufacturer Narrative
|
|
Conclusion: according to the information received, the finetuner echo started to smell burned and the user's mother got a small electric shock.During device investigation a failed capacitor was detected which was clearly the reason for the reported issue.Electrical inspection could not be performed because of the observed malfunction.No injury was reported.This is a final report.
|
| |
|