| Model Number |
N/A |
| Medical Device Problem Codes |
Break (1069); Thermal Decomposition of Device (1071)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This complaint is recorded by zimmer biomet under (b)(4).A follow up/ final report will be submitted once investigation is complete.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
It was reported that the ferrite inside has broken that caused a short circuit as there is a burn mark on the electronic board.There was no harm or delay reported.No adverse events were reported as a result of this malfunction.Follow up is in process as of today no additional information has been received.
|
| |
|
Event or Problem Description
|
|
The event timing was outside of surgery.
|
| |
|
Event or Problem Description
|
|
There is no additional event information available.
|
| |
|
Additional Manufacturer Narrative
|
|
Review of the most recent repair record could not be completed because the device was not returned for evaluation and repair.G2: foreign country : france.Review of the device history record identified no deviations or anomalies during manufacturing.A definitive root cause cannot be determined.The event cannot be confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|