• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MENTOR TEXAS MENTOR MEMORYSHAPE BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MENTOR TEXAS MENTOR MEMORYSHAPE BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED Back to Search Results
Model Number 3341359
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Codes Capsular Contracture (1761); Deformity/ Disfigurement (2360)
Date of Event 02/23/2023
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a 60-year-old caucasian female patient underwent revision breast augmentation with 495cc mentor memoryshape breast implant and experienced breast implant rupture on her right side postoperatively; diagnosed via mammogram and ultrasound scans.The patient has consulted with the healthcare professional.As a result, the devices will be explanted on (b)(6), 2023.
 
Additional Manufacturer Narrative
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: right rupture.Explantation date: (b)(6), 2023 mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
On march 15, 2023, the replacement 495 tm+ implants were ordered for patient's replacement surgery on march 23, 2023.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
On march 28, 2023, mentor became aware that the patient also experienced bilateral capsular contracture; baker grade unknown in addition to right side rupture.Clinical code "capsular contracture" has been added under field h6 on this form.Reason for device explant and/or reoperation: right rupture, and capsular contracture; baker grade unknown.This supplemental medwatch report is for the patient¿s right-sided device.Left side complication is being reported separately.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
On april 21, 2023, device evaluation was completed.Device evaluation summary: mentor conducted a visual inspection and microscopic examination of the returned device.Visual analysis of the returned sample revealed that the mentor memoryshape¿ tm+ 495cc breast implant was found to be ruptured.A microscopic examination was performed, and the cause of the tear could not be identified.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.Although no conclusion could be reached on the cause of the reported event, the instructions for use contain the following caution: rupture can occur at any time after implantation, but it is more likely to occur the longer the implant is implanted.The following things may cause the implant to rupture: damage by surgical instruments; stressing the implant during implantation and weakening it; folding or wrinkling of the implant shell; excessive force to the chest (e.G.During closed capsulotomy); trauma; compression during mammographic imaging; and severe capsular contracture.Breast implants may also simply wear out over time.Mentor concluded that the capsular contracture in the patient´s breast was the result of the body´s individual physiological response to the implantation of a foreign object in soft tissue.Capsular contracture is a known complication associated with these devices and is referenced in our current product insert data sheet.The american society of plastic surgeons recommends and encourages member surgeons to always submit breast implants, capsules, and effusion to pathology for examination.As part of mentor¿s quality process, all devices are manufactured, inspected, and released to approved specifications.Capa activity was not required.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.This supplemental medwatch report is for the patient¿s right-sided device.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
On april 10, 2023, the mentor failure analysis lab received the device for evaluation.The analysis has begun, but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.This supplemental medwatch report is for the patient¿s right-sided device.Manufacturer¿s reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENTOR MEMORYSHAPE BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer (Section G)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer Contact
kate karberg
3041 skyway circle north
irving, TX 75038
3035526892
MDR Report Key16530311
Report Number1645337-2023-02805
Device Sequence Number14108780
Product Code FTR
UDI-Device Identifier00081317005100
UDI-Public00081317005100
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
P060028
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/18/2020
Device Model Number3341359
Device Catalogue Number3341359
Device Lot Number6913403
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 02/27/2023
Supplement Date Received by Manufacturer03/15/2023
03/28/2023
04/10/2023
04/21/2023
Initial Report FDA Received Date03/13/2023
Supplement Report FDA Received Date03/28/2023
04/05/2023
04/19/2023
04/21/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/12/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age60 YR
Patient SexFemale
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-