| Model Number |
OPT318 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Fisher & paykel healthcare (f&p) are currently in the process of retrieving further information regarding this complaint.We will provide a follow up report upon the completion of our investigation.
|
| |
|
Event or Problem Description
|
|
A healthcare facility in washington reported via a fisher and paykel (f&p) healthcare field representative that the tubing of an opt318 optiflow junior cannula was detached from the swivel grip joint during use on a patient.There was no reported patient consequence.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).The optiflow junior nasal cannula is designed specifically for the delicate anatomical features and flow requirements of neonatal and paediatric patients.It features adhesive pads (wigglepads) to maintain cannula stability on the patient's cheeks, soft-touch nasal prongs and breathable kink-proof and crush-resistant flexible tubing.Method: the complaint opt318 optiflow junior nasal cannula was not returned to fisher & paykel healthcare (f&p) for evaluation.Our investigation is thus based on the pictures and information provided by the customer and our knowledge of the product.Results: visual inspection of the provided pictures revealed that one of the tubes detached from the swivel grip joint and the other tube was still attached to the joint, but it was torn.Conclusion: without the complaint device, we are unable to determine the cause of the reported event.All optiflow junior cannula are 100% leak and occlusion tested after final assembly and any cannula that fails is discarded.In addition, samples are taken hourly from each run and pull tested to check glue joint strength at the cannula/tube joint, as well as the swivel grip joint.If there are any failures the entire batch is put aside for further investigation.The subject opt316 optiflow junior nasal cannula would have met the required specification at the time of production.The user instructions which accompany the opt316 optiflow junior nasal cannula show in pictorial format the correct placement and fitting of the cannula and also warn: ensure that all connections are secure during use.Check cannula is undamaged and that the flow path is maintained.Under excessive load, the cannula may disconnect to prevent forces being transferred to the patient.Appropriate patient monitoring must be used at all times.Failure to monitor the patient may result in loss of therapy, serious injury or death.Do not stretch or crush tube.
|
| |
|
Event or Problem Description
|
|
A healthcare facility in washington reported via a fisher and paykel (f&p) healthcare field representative that the tubing of an opt318 optiflow junior cannula was detached from the swivel grip joint during use on a patient.There was no reported patient consequence.
|
| |
|