• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER MEDICAL-KALAMAZOO ISOLIBRIUM SUPPORT SURFACE - D; BED, PATIENT ROTATION, POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER MEDICAL-KALAMAZOO ISOLIBRIUM SUPPORT SURFACE - D; BED, PATIENT ROTATION, POWERED Back to Search Results
Model Number 2972
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Pressure Sore/Ulcer (2326)
Date of Event 02/13/2023
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient sustained a type 2 pressure injury.
 
Event or Problem Description
It was reported that the patient sustained a type 2 pressure injury.
 
Additional Manufacturer Narrative
The surface was evaluated by a stryker field service technician, who found that the hospital acquired pressure injury was not due to any component level defect or malfunction of the surface.Additionally, a stryker quality assurance engineer spoke with the sales account manager who explained that the mattress had lost power from the bed.The nurse attending to the patient was new and did not check the foot end of the bed to ensure the mattress was powered and detected by the bed.It is unknown how long the surface was without power.The patient received a stage 2 pressure injury.The issue was resolved for the customer by the sales rep conducting additional training with the nurse and confirming no further action is needed by stryker at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ISOLIBRIUM SUPPORT SURFACE - D
Common Device Name
BED, PATIENT ROTATION, POWERED
Manufacturer (Section D)
STRYKER MEDICAL-KALAMAZOO
3800 east centre avenue
portage MI 49002
Manufacturer (Section G)
STRYKER MEDICAL-KALAMAZOO
3800 east centre avenue
portage MI 49002
Manufacturer Contact
marleca lane
3800 east centre avenue
portage, MI 49002
2693292100
MDR Report Key16531252
Report Number0001831750-2023-00311
Device Sequence Number6126733
Product Code IKZ
UDI-Device Identifier07613327278453
UDI-Public07613327278453
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Model Number2972
Device Catalogue Number2972000000
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/13/2023
Supplement Date Received by Manufacturer02/13/2023
Initial Report FDA Received Date03/13/2023
Supplement Report FDA Received Date04/06/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/30/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-