| Model Number |
A2025-120 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/22/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that a 2.5mmx120mm armada 14 balloon dilatation catheter (bdc) ruptured at 6 atmospheres (atm).No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Date of event: estimated.Manufacturer's investigation is still pending at this time.Results and conclusions will be provided in the final report.
|
| |
|
Additional Manufacturer Narrative
|
|
Visual inspection was performed on the returned device.The reported balloon rupture was confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot.The investigation was unable to determine a conclusive cause for the reported complaint.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.
|
| |
|