| Model Number |
1010 |
| Medical Device Problem Codes |
Incorrect, Inadequate or Imprecise Result or Readings (1535); Noise, Audible (3273)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/30/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that this device exhibited fault codes for long charge times.The device has been implanted greater than seven years and has 5% battery remaining.A review of device downloaded memory was done by technical services and the long charge times were confirmed.The device remains implanted, however technical services recommended explant.There were no adverse patient effects.
|
| |
|
Additional Manufacturer Narrative
|
|
Upon receipt at our post market quality assurance laboratory, the device was examined.Visual inspection noted no damage.The device was then exposed to simulated heart load conditions, and the defibrillation functions were tested.The device operated appropriately, according to its performance specifications with no interruptions in therapy output or programmer communication at the returned programmed settings.A series of electrical tests was also performed, and again, normal device function was observed.Laboratory analysis did not identify any characteristics that would have caused or contributed to the reported clinical observations.
|
| |
|
Event or Problem Description
|
|
This supplemental report is being filed to provide additional information regarding product analysis.This supplemental report is being filed to provide additional information that the product has been explanted and replaced.There were no additional adverse patient effects.It was reported that this device exhibited fault codes for long charge times.The device has been implanted greater than seven years and has 5% battery remaining.A review of device downloaded memory was done by technical services and the long charge times were confirmed.The device remains implanted, however technical services recommended explant.There were no adverse patient effects.
|
| |
|
Event or Problem Description
|
|
This supplemental report is being filed to provide additional information that the product has been explanted and replaced.There were no additional adverse patient effects.It was reported that this device exhibited fault codes for long charge times.The device has been implanted greater than seven years and has 5% battery remaining.A review of device downloaded memory was done by technical services and the long charge times were confirmed.The device remains implanted, however technical services recommended explant.There were no adverse patient effects.
|
| |
|